Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy

Authors
 Seunghyeop Baek  ;  Jin Sung Kim  ;  Woo Sang Ahn  ;  Sohyun Ahn 
Citation
 JOURNAL OF MAGNETICS, Vol.25(4) : 614-619, 2020-12 
Journal Title
JOURNAL OF MAGNETICS
ISSN
 1226-1750 
Issue Date
2020-12
Keywords
electromagnetic spectrum ; CZT detector ; dual-energy CBCT ; energy modulation filter ; energy in electro magnetic radiation
Abstract
To implement dual-energy cone-beam computed tomography (CBCT), a 1 to 3 mm thick copper filter was
applied in front of the kilovolt generator mounted on a medical linear accelerator. The energy calibration was
performed from three energy peaks among the electromagnetic radiation, gamma rays, emitted by the 192Ir
source. The energy spectrum was measured using a cadmium zinc telluride (CZT) detector. A figure-of-merit
(FOM) value was defined and evaluated in order to consider both the increase in the mean energy and the
decrease in the intensity according to the filter. For a 1 mm thick filter, the FOM values were 0.82 and 0.92 for
tube voltages of 100 and 120 kVp, respectively. The intens ity was maintained reasonably and the effect of
increasing the mean energy was evident, thus indicating its suitability for implementing dual-energy CBCT.
Files in This Item:
T999202157.pdf Download
DOI
10.4283/JMAG.2020.25.4.614
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Humanities and Social Sciences (인문사회의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, So Yoon(김소윤) ORCID logo https://orcid.org/0000-0001-7015-357X
Kim, Jinsung(김진성) ORCID logo https://orcid.org/0000-0003-1415-6471
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/184926
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links