0 539

Cited 21 times in

Analysis of 30 Breast Implant Rupture Cases

Authors
 Kwan Chul Tark  ;  Hii Sun Jeong  ;  Tae Suk Roh  ;  Jong Woo Choi 
Citation
 AESTHETIC PLASTIC SURGERY, Vol.29(6) : 460-469, 2005 
Journal Title
AESTHETIC PLASTIC SURGERY
ISSN
 0364-216X 
Issue Date
2005
MeSH
Adult ; Breast Implants* ; Female ; Humans ; Middle Aged ; Postoperative Complications ; Reoperation ; Retrospective Studies
Keywords
MRI ; Radiologic sign ; Rupture of breast implant
Abstract
Breast implants used for augmentation mammoplasty or breast reconstruction could rupture from various causes such as trauma or spontaneous failure. The objectives of this study were to investigate the relationships between the causes of implant rupture and the degree of capsular contracture, and then to evaluate the relative efficacies of specific signs on magnetic resonance imaging (MRI) known to be beneficial for diagnosing the rupture. A retrospective review identified patients with prosthetic implant rupture or impending rupture treated by the senior author. The 30 cases of implant rupture available for review were classified into two groups: intracapsular and extracapsular ruptures. The 30 cases of breast implant ruptures were analyzed with respect to the clinical symptoms and signs, the causes of rupture, the degree of capsular contracture, and therapeutic plans. Among the 30 cases, 14 patients who had undergone MRI during the diagnostic period were analyzed with respect to the relationships between MRI readings and operative findings. Spontaneous rupture of membranes was most common (80%), followed by failure because of trauma (7%) and valve or implant base (4%). The symptoms during implant rupture were contour deformity, palpated masslike lesions, pain, and focal inflammation. According to the analysis of specific MRI signs, the sensitivity and specificity of the linguine sign were 87% and 100%, respectively, for intracapsular rupture. For extracapsular rupture, the sensitivity and specificity of the linguine sign were, respectively, 67% and 75%. The sensitivity and specificity of the rat-tail sign and tear drop sign were 14% and 50%, respectively.
Breast implant rupture was correlated with the degree of capsular contracture in our study. Among the various specific MRI signs used in diagnosing the rupture, the linguine sign was reliable and had a high sensitivity and specificity, especially in cases of intracapsular rupture. On the other hand, the rat-tail and tear drop signs were nonspecific signs for diagnosing the rupture of breast implant.
Full Text
http://link.springer.com/article/10.1007%2Fs00266-004-0146-x
DOI
10.1007/s00266-004-0146-x
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Plastic and Reconstructive Surgery (성형외과학교실) > 1. Journal Papers
Yonsei Authors
Roh, Tai Suk(노태석) ORCID logo https://orcid.org/0000-0001-8681-159X
Jeong, Hii Sun(정희선)
Choi, Jong Woo(최종우)
Tark, Kwan Chul(탁관철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/151332
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links