Des-γ-carboxyprothrombin (DCP) ; Hepatocellular carcinoma (HCC) ; Protein induced by vitamin K absence or antagonist-II (PIVKA-II) ; Automated analyzer ; Reference range
Abstract
OBJECTIVES: We evaluated two new autoanalyzers, μTAS and Lumipulse for des-γ-carboxyprothrombin (DCP) assay.
MATERIAL AND METHODS: Analytical performance was evaluated, and the upper reference limit of the 97.5th percentile for DCP was re-established using sera from 140 healthy individuals. DCP levels were determined by the two autoanalyzers and EIA in a total of 239 sera from HCC patients (n=120) and those without HCC (n=119).
RESULTS: Total imprecision of the two automated assays was <5% CV. Analytical measurement ranges (AMRs) were verified to be linear. The new reference limits were 29.5 mAU/mL for μTAS and 35.0 mAU/mL for Lumipulse. There were proportional and constant biases between the results from the autoanalyzers and those from EIA.