103 79

Cited 9 times in

Evaluation of a fully automated treponemal test and comparison with conventional VDRL and FTA-ABS tests.

Authors
 Yongjung Park ; Younhee Park ; Hyon-Suk Kim ; Myoung Hee Park ; Shin Young Joo 
Citation
 American Journal of Clinical Pathology, Vol.136(5) : 705~710, 2011 
Journal Title
 American Journal of Clinical Pathology 
ISSN
 0002-9173 
Issue Date
2011
Abstract
We evaluated analytic performances of an automated treponemal test and compared this test with the Venereal Disease Research Laboratory test (VDRL) and fluorescent treponemal antibody absorption test (FTA-ABS). Precision performance of the Architect Syphilis TP assay (TP; Abbott Japan, Tokyo, Japan) was assessed, and 150 serum samples were assayed with the TP before and after heat inactivation to estimate the effect of heat inactivation. A total of 616 specimens were tested with the FTA-ABS and TP, and 400 were examined with the VDRL. The TP showed good precision performance with total imprecision of less than a 10% coefficient of variation. An excellent linear relationship between results before and after heat inactivation was observed (R(2) = 0.9961). The FTA-ABS and TP agreed well with a κ coefficient of 0.981. The concordance rate between the FTA-ABS and TP was the highest (99.0%), followed by the rates between FTA-ABS and VDRL (85.0%) and between TP and VDRL (83.8%). The automated TP assay may be adequate for screening for syphilis in a large volume of samples and can be an alternative to FTA-ABS.
URI
http://ir.ymlib.yonsei.ac.kr/handle/22282913/94556
DOI
10.1309/AJCP0WOK0QDYGDCM
Appears in Collections:
1. 연구논문 > 1. College of Medicine > Dept. of Laboratory Medicine
Yonsei Authors
사서에게 알리기
  feedback
Files in This Item:
T201103922.pdfDownload
Export
RIS (EndNote)
XLS (Excel)
XML

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.