116 151

Cited 2 times in

Myocardial T2* Imaging at 3T and 1.5T: A Pilot Study with Phantom and Normal Myocardium

Authors
 Suyon Chang  ;  Jinho Park  ;  Young-Joong Yang  ;  Kyongmin Sarah Beck  ;  Pan Ki Kim  ;  Byoung Wook Choi  ;  Jung Im Jung 
Citation
 JOURNAL OF CARDIOVASCULAR DEVELOPMENT AND DISEASE, Vol.9(8) : 271, 2022-08 
Journal Title
JOURNAL OF CARDIOVASCULAR DEVELOPMENT AND DISEASE
Issue Date
2022-08
Keywords
3T ; T2* ; heart ; iron overload ; magnetic resonance imaging
Abstract
Background: Myocardial T2* mapping at 1.5T remains the gold standard, but the use of 3T scanners is increasing. We aimed to determine the conversion equations in different scanners with clinically available, vendor-provided T2* mapping sequences using a phantom and evaluated the feasibility of the phantom-based conversion method.

Methods: T2* of a phantom with FeCl3 (five samples, 3.53-20.09 mM) were measured with 1.5T (MR-A1) and 3T scanners (MR-A2, A3, B), and the site-specific equation was determined. T2* was measured in the interventricular septum of three healthy volunteers at 1.5T (T2*1.5T, MR-A1) and 3T (T2*3.0T, MR-B). T2*3.0T was converted based on the equation derived from the phantom (T2*eq).

Results: R2* at 1.5T and 3T showed linear association, but a different relationship was observed according to the scanners (MR-A2, R2*1.5T = 0.76 × R2*3.0T - 2.23, R2 = 0.999; MR-A3, R2*1.5T = 0.95 × R2*3.0T - 34.28, R2 = 0.973; MR-B, R2*1.5T = 0.76 × R2*3.0T - 3.02, R2 = 0.999). In the normal myocardium, T2*eq and T2*1.5T showed no significant difference (35.5 ± 3.5 vs. 34.5 ± 1.2, p = 0.340). The mean squared error between T2*eq and T2*1.5T was 16.33, and Bland-Altman plots revealed a small bias (-0.94, 95% limits of agreement: -8.86-6.99).

Conclusions: a phantom-based, site-specific equation can be utilized to estimate T2* values at 1.5T in centers where only 3T scanners are available.
Files in This Item:
T9992022657.pdf Download
DOI
10.3390/jcdd9080271
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Choi, Byoung Wook(최병욱) ORCID logo https://orcid.org/0000-0002-8873-5444