124 280

Cited 5 times in

Automated Measurement of Native T1 and Extracellular Volume Fraction in Cardiac Magnetic Resonance Imaging Using a Commercially Available Deep Learning Algorithm

Authors
 Suyon Chang  ;  Kyunghwa Han  ;  Suji Lee  ;  Young Joong Yang  ;  Pan Ki Kim  ;  Byoung Wook Choi  ;  Young Joo Suh 
Citation
 KOREAN JOURNAL OF RADIOLOGY, Vol.23(12) : 1251-1259, 2022-12 
Journal Title
KOREAN JOURNAL OF RADIOLOGY
ISSN
 1229-6929 
Issue Date
2022-12
MeSH
Algorithms ; Deep Learning* ; Heart ; Humans ; Magnetic Resonance Imaging ; Myocardium
Keywords
Deep learning ; Extracellular volume fraction ; Heart ; Magnetic resonance imaging ; T1 mapping
Abstract
Objective: T1 mapping provides valuable information regarding cardiomyopathies. Manual drawing is time consuming and prone to subjective errors. Therefore, this study aimed to test a DL algorithm for the automated measurement of native T1 and extracellular volume (ECV) fractions in cardiac magnetic resonance (CMR) imaging with a temporally separated dataset.

Materials and methods: CMR images obtained for 95 participants (mean age ± standard deviation, 54.5 ± 15.2 years), including 36 left ventricular hypertrophy (12 hypertrophic cardiomyopathy, 12 Fabry disease, and 12 amyloidosis), 32 dilated cardiomyopathy, and 27 healthy volunteers, were included. A commercial deep learning (DL) algorithm based on 2D U-net (Myomics-T1 software, version 1.0.0) was used for the automated analysis of T1 maps. Four radiologists, as study readers, performed manual analysis. The reference standard was the consensus result of the manual analysis by two additional expert readers. The segmentation performance of the DL algorithm and the correlation and agreement between the automated measurement and the reference standard were assessed. Interobserver agreement among the four radiologists was analyzed.

Results: DL successfully segmented the myocardium in 99.3% of slices in the native T1 map and 89.8% of slices in the post-T1 map with Dice similarity coefficients of 0.86 ± 0.05 and 0.74 ± 0.17, respectively. Native T1 and ECV showed strong correlation and agreement between DL and the reference: for T1, r = 0.967 (95% confidence interval [CI], 0.951-0.978) and bias of 9.5 msec (95% limits of agreement [LOA], -23.6-42.6 msec); for ECV, r = 0.987 (95% CI, 0.980-0.991) and bias of 0.7% (95% LOA, -2.8%-4.2%) on per-subject basis. Agreements between DL and each of the four radiologists were excellent (intraclass correlation coefficient [ICC] of 0.98-0.99 for both native T1 and ECV), comparable to the pairwise agreement between the radiologists (ICC of 0.97-1.00 and 0.99-1.00 for native T1 and ECV, respectively).

Conclusion: The DL algorithm allowed automated T1 and ECV measurements comparable to those of radiologists.
Files in This Item:
T202205885.pdf Download
DOI
10.3348/kjr.2022.0496
Appears in Collections:
1. College of Medicine (의과대학) > Research Institute (부설연구소) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Suh, Young Joo(서영주) ORCID logo https://orcid.org/0000-0002-2078-5832
Lee, Suji(이수지) ORCID logo https://orcid.org/0000-0002-8770-622X
Choi, Byoung Wook(최병욱) ORCID logo https://orcid.org/0000-0002-8873-5444
Han, Kyung Hwa(한경화)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/192836
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links