178 301

Cited 33 times in

Comparison Between an Intraocular Lens With Extended Depth of Focus (Tecnis Symfony ZXR00) and a New Monofocal Intraocular Lens With Enhanced Intermediate Vision (Tecnis Eyhance ICB00)

Authors
 Young Joon Jeon  ;  Yisang Yoon  ;  Tae-Im Kim  ;  Kyungmin Koh 
Citation
 ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY, Vol.10(6) : 542-547, 2021-11 
Journal Title
ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY
Issue Date
2021-11
MeSH
Aged ; Humans ; Lens Implantation, Intraocular ; Lenses, Intraocular* ; Middle Aged ; Patient Satisfaction ; Phacoemulsification* ; Prospective Studies ; Prosthesis Design
Abstract
Purpose: This study compared the extended depth of focus (EDOF) intraocular lens (IOL) (ZXR00; Tecnis Symfony, Johnson & Johnson Vision, Santa Ana, CA, US) to a novel, higher-order aspheric monofocal IOL (ICB00; Tecnis Eyhance, Johnson & Johnson Vision, Santa Ana, CA, US) which uses the same platform and material.

Methods: Medical records of patients undergoing cataract surgery with ZXR00 or ICB00 implantation between March 2020 and January 2021 and with the data available for the 3-month visit were reviewed. The uncorrected near, intermediate, and distance visual acuity (VA); corrected distance VA; and optical quality parameters were the main outcome measures.

Results: Among the 174 enrolled patients, 72 and 102 received the ZXR00 and ICB00, respectively. The average patient ages were 59.6 ± 10.6 (range: 49 to 70) and 65.2 ± 8.2 (range: 45 to 82) years in the ZXR00 and ICB00 groups, respectively, with significantly older patients in the ICB00 group. The other baseline parameters were not different for the 2 groups. Compared to the ICB00 group, the ZXR00 group showed markedly superior near VA (P < 0.05) at 3 months postoperatively. In terms of optical quality, ICB00 was, statistically, significantly superior to ZXR00.

Conclusions: The ZXR00 showed remarkable near vision and defocus curve smoothness, while the ICB00 achieved better optical quality. The 2 IOLs had comparable distance and intermediate vision.
Files in This Item:
T202126201.pdf Download
DOI
10.1097/APO.0000000000000439
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/190600
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.