Accuracy of Continuous and Noninvasive Hemoglobin Monitoring in the Presence of CO2 Insufflation: An Observational Pilot Study

Authors
 Hye Sun Lee  ;  Ji Won Yoo  ;  Ha Yeon Kim  ;  Na Young Kim  ;  Ji Eun Kim 
Citation
 MEDICAL SCIENCE MONITOR, Vol.27 : e933027, 2021-10 
Journal Title
MEDICAL SCIENCE MONITOR
ISSN
 1234-1010 
Issue Date
2021-10
MeSH
Aged ; Carbon Dioxide / metabolism ; Female ; Hemoglobins / analysis* ; Humans ; Insufflation ; Laparoscopy / methods* ; Male ; Middle Aged ; Monitoring, Intraoperative / methods* ; Pilot Projects
Abstract
BACKGROUND Laparoscopic surgery has several benefits, but it requires prolonged carbon dioxide (CO₂) insufflation. Several factors affect the accuracy of continuous and noninvasive hemoglobin (SpHb) monitoring, but the effects of CO₂ insufflation are undetermined. This study investigated the effect of CO₂ insufflation on SpHb monitoring in laparoscopic surgery. MATERIAL AND METHODS Twenty patients undergoing laparoscopic gastrectomy were enrolled. Anesthesia was maintained using sevoflurane and remifentanil within an end-tidal CO₂ of 30-45 mmHg. The CO₂ insufflation was maintained at 12 mmHg using CO₂. SpHb was monitored with a Radical-7 Pulse CO-Oximeter, and laboratory hemoglobin (tHb) was analyzed using a satellite blood analyzer. RESULTS Forty paired measurements were analyzed. The mean perfusion index, SpHb, and tHb were 3.10±1.77%, 10.92±1.48 g/dL, and 11.51±0.88 g/dL, respectively. SpHb underestimated tHb with a bias (precision) of -0.59 (1.28 g/dL), and the 95% limit of agreement was wide (-3.11 to 1.92 g/dL). SpHb was moderately correlated with tHb (r=0.50, 95% CI: 0.23 to 0.70). The concordance rate was 67%. ΔSpHb was not correlated with ΔtHb (r=0.29, 95% CI: -0.18 to -0.65). A similar bias, wider limits of agreement, a higher
SpHb-tHb
, but more significant correlation between SpHb and tHb were observed for the "PaCO2 <40 mmHg" range compared with the "40 mmHg ≤PaCO₂" range. CONCLUSIONS SpHb may have an acceptable accuracy but has a weak trending ability in the presence of CO₂ insufflation, and it can be affected by PaCO₂. Further research on the effects of CO₂ insufflation on SpHb is needed.
Files in This Item:
T202125781.pdf Download
DOI
10.12659/MSM.933027
Appears in Collections:
1. College of Medicine (의과대학) > Yonsei Biomedical Research Center (연세의생명연구원) > 1. Journal Papers
Yonsei Authors
Lee, Hye Sun(이혜선) ORCID logo https://orcid.org/0000-0001-6328-6948
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/188183
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links