Abstract
Background: The relationship between the characteristics of cardiac implantable electronic device (CIED) leads and subclinical cardiac perforations remains unclear. This study aimed to evaluate the incidence of subclinical cardiac perforation among various CIED leads using cardiac computed tomography (CT).

Methods: A total of 271 consecutive patients with 463 CIED leads, who underwent cardiac CT after CIED implantation, were included in this retrospective observational study. Cardiac CT images were reviewed by one radiologist and two cardiologists. Subclinical perforation was defined as traversal of the lead tip past the outer myocardial layer without symptoms and signs related to cardiac perforation. We compared the subclinical cardiac perforation rates of the available lead types.

Results: A total of 219, 49, and 3 patients had pacemakers, implantable cardioverter-defibrillators, and cardiac resynchronization therapy, respectively. The total subclinical cardiac perforation rate was 5.6%. Subclinical cardiac perforation by screw-in ventricular leads was significantly more frequent than that caused by tined ventricular leads (13.3% vs 3.3%, respectively, p = 0.002). There were no significant differences in the incidence of cardiac perforation between atrial and ventricular leads, screw-in and tined atrial leads, pacing and defibrillator ventricular leads, nor between magnetic resonance (MR)-conditional and MR-unsafe screw-in ventricular leads. Screw-in ventricular leads were significantly associated with subclinical cardiac perforation [odds ratio, 4.554; 95% confidence interval, 1.587-13.065, p = 0.005]. There was no case subclinical cardiac perforation by septal ventricular leads.

Conclusions: Subclinical cardiac perforation by screw-in ventricular leads is not rare. Septal pacing may be helpful in avoiding cardiac perforation.
Files in This Item:
T202103375.pdf Download
DOI
10.1186/s12872-021-02159-3
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Min(김민) ORCID logo https://orcid.org/0000-0002-8132-9873
Kim, Young Jin(김영진) ORCID logo https://orcid.org/0000-0002-6235-6550
Kim, In-Soo(김인수) ORCID logo https://orcid.org/0000-0003-2801-5514
Kim, Tae-Hoon(김태훈) ORCID logo https://orcid.org/0000-0003-4200-3456
Pak, Hui Nam(박희남) ORCID logo https://orcid.org/0000-0002-3256-3620
Uhm, Jae Sun(엄재선) ORCID logo https://orcid.org/0000-0002-1611-8172
Yu, Hee Tae(유희태) ORCID logo https://orcid.org/0000-0002-6835-4759
Lee, Moon-Hyoung(이문형) ORCID logo https://orcid.org/0000-0002-7268-0741
Lee, Hye Jeong(이혜정) ORCID logo https://orcid.org/0000-0003-4349-9174
Joung, Bo Young(정보영) ORCID logo https://orcid.org/0000-0001-9036-7225
Jin, Moo Nyun(진무년) ORCID logo https://orcid.org/0000-0001-5482-4441
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/184668
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links