Prediction accuracy of conventional and total keratometry for intraocular lens power calculation in femtosecond laser-assisted cataract surgery

Authors
 Soyoung Ryu  ;  Ikhyun Jun  ;  Tae-Im Kim  ;  Kyoung Yul Seo  ;  Eung Kweon Kim 
Citation
 SCIENTIFIC REPORTS, Vol.11(1) : 12869, 2021-06 
Journal Title
SCIENTIFIC REPORTS
Issue Date
2021-06
Abstract
This study evaluated the accuracy of total keratometry (TK) and standard keratometry (K) for intraocular lens (IOL) power calculation in eyes treated with femtosecond laser-assisted cataract surgery. The retrospective study included a retrospective analysis of data from 62 patients (91 eyes) who underwent uneventful femtosecond laser-assisted cataract surgery with Artis PL E (Cristalens Industrie, Lannion, France) IOL implantation by a single surgeon between May 2020 and December 2020 in Severance Hospital, Seoul, South Korea. The new IOLMaster 700 biometry device (Carl Zeiss Meditec, Jena, Germany) was used to calculate TK and K. The mean absolute error (MAE), median absolute error (MedAE), and the percentages of eyes within prediction errors of ± 0.25 D, ± 0.50 D, and ± 1.00 D were calculated for all IOL formulas (SRK/T, Hoffer-Q, Haigis, Holladay 1, Holladay 2, and Barrett Universal II). There was strong agreement between K and TK (intraclass correlation coefficient = 0.99), with a mean difference of 0.04 D. For all formulas, MAE tended to be lower for TK than for K, and relatively lower MAE and MedAE values were observed for SRK/T and Holladay 1. Furthermore, for all formulas, a greater proportion of eyes fell within ± 0.25 D of the predicted postoperative spherical equivalent range in the TK group than in the K group. However, differences in MAEs, MedAEs, and percentages of eyes within the above prediction errors were not statistically significant. In conclusion, TK and K exhibit comparable performance for refractive prediction in eyes undergoing femtosecond laser-assisted cataract surgery.
Files in This Item:
T202102556.pdf Download
DOI
10.1038/s41598-021-92354-1
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Jun, Ik Hyun(전익현) ORCID logo https://orcid.org/0000-0002-2160-1679
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/184199
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links