Monitoring beam-quality constancy considering uncertainties associated with ionization chambers in Daily QA3 device

Authors
 Su Chul Han  ;  Jihun Kim  ;  Min Cheol Han  ;  Kyung Hwan Chang  ;  Kwangwoo Park  ;  Ho Jin Kim  ;  Dong Wook Kim  ;  Jin Sung Kim 
Citation
 PLOS ONE, Vol.16(2) : e0246845, 2021-02 
Journal Title
PLOS ONE
Issue Date
2021-02
Abstract
This study evaluates the changes occurring in the X-ray energy of a linear accelerator (LINAC) using a Daily QA3 detector system. This is accomplished by comparing the Daily QA3 results against those obtained using a water phantom. The X-energy levels of a LINAC were monitored over a duration of 1 month using the Daily QA3 system. Moreover, to account for the uncertainty, the reproducibility of the Daily QA3 ionization-chamber results was assessed by performing repeated measurements (12 per day). Subsequently, the energy-monitoring results were compared with the energy-change results calculated using the water-phantom percentage depth dose (PDD) ratio. As observed, the 6- and 10-MV beams experienced average daily energy-level changes of (-0.30 ± 0.32)% and (0.05 ± 0.38)%, respectively, during repeated measurements. The corresponding energy changes equaled (-0.30 ± 0.55)% and (-0.05 ± 0.48)%, respectively, when considering the measurement uncertainty. The Daily QA3 measurements performed at 6 MV demonstrated a variation of (2.15 ± 0.81)% (i.e., up to 3%). Meanwhile, the corresponding measurements performed using a water phantom demonstrated an increase in the PDD ratio from 0.577 to 0.580 (i.e., approximately 0.5%). At 10 MV, the energy variation in the Daily QA3 measurements equaled (-0.41 ± 0.82)% (i.e., within 1.5%), whereas the corresponding water phantom PDD ratio remained constant at 0.626. These results reveal that the Daily QA3 system can be used to monitor small energy changes occurring within radiotherapy machines. This demonstrates its potential for use as a secondary system for monitoring energy changes as part of the daily quality-assurance workflow.
Files in This Item:
T202100536.pdf Download
DOI
10.1371/journal.pone.0246845
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, Dong Wook(김동욱) ORCID logo https://orcid.org/0000-0002-5819-9783
Kim, Jihun(김지훈) ORCID logo https://orcid.org/0000-0003-4856-6305
Kim, Jinsung(김진성) ORCID logo https://orcid.org/0000-0003-1415-6471
Kim, Hojin(김호진) ORCID logo https://orcid.org/0000-0002-4652-8682
Park, Kwang Woo(박광우) ORCID logo https://orcid.org/0000-0002-9843-7985
Chang, Kyung Hwan(장경환)
Han, Min Cheol(한민철)
Han, Su Chul(한수철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/182174
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links