362 349

Cited 0 times in

광학문자인식(OCR)에 의한 CT 방사선 선량 관리시스템 개발에 관한 연구

Other Titles
 The Study on Radiation Dose Monitoring System Development of Computed Tomography (CT) by Optical Character Recognition (OCR) 
Authors
 윤영욱  ;  김남현  ;  백창구 
Citation
 Journal of the Korean Society of Imaging Informatics in Medicine (대한의학영상정보학회지), Vol.18(1) : 27-31, 2012-12 
Journal Title
Journal of the Korean Society of Imaging Informatics in Medicine(대한의학영상정보학회지)
Issue Date
2012-12
Keywords
Dose monitoring system ; Dose report ; DICOM structured report ; Diagnostic reference report ; Optical c
Abstract
Since Korea Food and Drug Administration and The Ministry of Health-Welfare published guidelinefor its recommendations on issues related to patient radiation dose from CT X-ray to reduce patientdose in 2009, it has been followed up. As a result of this effort, recently many hospital has been buildingup dose monitoring system to collect and manage radiation dose systematically. In case CT whichis installed and used in hospital does support dose report by structured report, it is easy to providedose report to dose monitoring system. However It is very hard to integrate radiation dose from CT installedor manufactured before DICOM SR requirements containing dose report was released, whichwas defined by DICOM Standard Committee into dose monitoring system so that this makes it difficultto comply with guideline for diagnostic reference level. Therefore this report contains the experienceon integrating the value of DLP (Dose Length product) from CT installed in radiology department ofuniversity hospital of Seoul into dose monitoring system by using Optical Character Recognition technology.
Files in This Item:
T201206055.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Engineering (의학공학교실) > 1. Journal Papers
Yonsei Authors
Kim, Nam Hyun(김남현)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/178546
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.