0 336

Cited 41 times in

An assembly-type master-slave catheter and guidewire driving system for vascular intervention

Authors
 Hyo-Jeong Cha  ;  Byung-Ju Yi  ;  Jong Yun Won 
Citation
 PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, Vol.231(1) : 69-79, 2017-01 
Journal Title
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE
ISSN
 0954-4119 
Issue Date
2017-01
Keywords
Vascular intervention ; catheter and guidewire driving system ; endovascular surgery ; haptics ; master–slave robotic system
Abstract
Current vascular intervention inevitably exposes a large amount of X-ray to both an operator and a patient during the procedure. The purpose of this study is to propose a new catheter driving system which assists the operator in aspects of less X-ray exposure and convenient user interface. For this, an assembly-type 4-degree-of-freedom master-slave system was designed and tested to verify the efficiency. First, current vascular intervention procedures are analyzed to develop a new robotic procedure that enables us to use conventional vascular intervention devices such as catheter and guidewire which are commercially available in the market. Some parts of the slave robot which contact the devices were designed to be easily assembled and dissembled from the main body of the slave robot for sterilization. A master robot is compactly designed to conduct insertion and rotational motion and is able to switch from the guidewire driving mode to the catheter driving mode or vice versa. A phantom resembling the human arteries was developed, and the master-slave robotic system is tested using the phantom. The contact force of the guidewire tip according to the shape of the arteries is measured and reflected to the user through the master robot during the phantom experiment. This system can drastically reduce radiation exposure by replacing human effort by a robotic system for high radiation exposure procedures. Also, benefits of the proposed robot system are low cost by employing currently available devices and easy human interface.
Full Text
https://journals.sagepub.com/doi/full/10.1177/0954411916679328
DOI
10.1177/0954411916679328
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Won, Jong Yun(원종윤) ORCID logo https://orcid.org/0000-0002-8237-5628
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/178364
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links