Abstract
OBJECTIVES: The aim of the present study was to investigate the adaptation and guide hole tolerance of metal sleeve-free computer-assisted implant surgical guides fabricated with 3D printers.

METHODS: An implant surgical guide for full-guided implant placement was designed with a total of eight different guide holes. Ten implant surgical guides (n=10) were fabricated from the same design with each of five in-office 3D printers (D1, FOR, ONE, PER, and ZEN) using compatible printing materials. Ten surgical guides fabricated by the manufacturer of the implant company were used as the control group (CON). The adaptation of the surgical guides was evaluated by the replica technique. The tolerance of the guide holes was evaluated by measuring the degree of diversion with guide drills.

RESULTS: CON and D1 showed superior internal adaptation with a gap distance of less than 1mm. The mean degree of diversion of the guide holes ranged from 3.45° for ZEN to 6.55° for PER. The tolerances of CON (4.70°) and D1 (4.50°) did not differ at the level of statistical significance at α=0.05.

SIGNIFICANCE: The characteristics of implant surgical guides were evaluated per se. None of the 3D printers fabricated superior implant surgical guides to those produced by the manufacturer with regard to the internal fit and guide tolerance. However, the potential for the routine clinical use of in-office 3D printers was demonstrated. Further studies are required to determine how the guide hole tolerance and the angular deviation between the preplanned and actual implant positions are related.
Full Text
https://www.sciencedirect.com/science/article/pii/S0109564118310546
DOI
10.1016/j.dental.2019.01.001
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Prosthodontics (보철과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jong Eun(김종은) ORCID logo https://orcid.org/0000-0002-7834-2524
Kim, Jee Hwan(김지환) ORCID logo https://orcid.org/0000-0002-0872-4906
Park, Ji-Man(박지만) ORCID logo https://orcid.org/0000-0003-0018-1166
Shim, June Sung(심준성) ORCID logo https://orcid.org/0000-0003-1428-0122
Oh, Kyung Chul(오경철) ORCID logo https://orcid.org/0000-0003-4584-2597
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/169939
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links