390 624

Cited 14 times in

Mechanical and biological complication rates of the modified lateral-screw-retained implant prosthesis in the posterior region: an alternative to the conventional Implant prosthetic system.

Authors
 Jae-Hong Lee  ;  Jong-Bin Lee  ;  Man-Yong Kim  ;  Joon-Ho Yoon  ;  Seong-Ho Choi  ;  Young-Taek Kim 
Citation
 JOURNAL OF ADVANCED PROSTHODONTICS, Vol.8(2) : 150-157, 2016 
Journal Title
JOURNAL OF ADVANCED PROSTHODONTICS
ISSN
 2005-7806 
Issue Date
2016
Keywords
Dental implant ; Dental prosthesis design ; Peri-implantitis ; Prosthesis failure ; Single-tooth implant
Abstract
PURPOSE: The modified lateral-screw-retained implant prosthesis (LSP) is designed to combine the advantages of screw- and cement-retained implant prostheses. This retrospective study evaluated the mechanical and biological complication rates of implant-supported single crowns (ISSCs) inserted with the modified LSP in the posterior region, and determined how these complication rates are affected by clinical factors.
MATERIALS AND METHODS: Mechanical complications (i.e., lateral screw loosening [LSL], abutment screw loosening, lateral screw fracture, and ceramic fracture) and biological complications (i.e., peri-implant mucositis [PM] and peri-implantitis) were identified from the patients' treatment records, clinical photographs, periapical radiographs, panoramic radiographs, and clinical indices. The correlations between complication rates and the following clinical factors were determined: gender, age, position in the jaw, placement location, functional duration, clinical crown-to-implant length ratio, crown height space, and the use of a submerged or nonsubmerged placement procedure.
RESULTS: Mechanical and biological complications were present in 25 of 73 ISSCs with the modified LSP. LSL (n=11) and PM (n=11) were the most common complications. The incidence of mechanical complications was significantly related to gender (P=.018). The other clinical factors were not significantly associated with mechanical and biological complication rates.
CONCLUSION: Within the limitations of this study, the incidence of mechanical and biological complications in the posterior region was similar for both modified LSP and conventional implant prosthetic systems. In addition, the modified LSP is amenable to maintenance care, which facilitates the prevention and treatment of mechanical and biological complications.
Files in This Item:
T201602063.pdf Download
DOI
10.4047/jap.2016.8.2.150
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Periodontics (치주과학교실) > 1. Journal Papers