520 705

Cited 0 times in

개인용 컴퓨터에서 재구성한 3차원 전산화단층영상의 두부계측 재현성

Other Titles
 Reproducibilities of cephalometric measurements of three-dimensional CT images reconstructed in the personal computer 
Authors
 전국진  ;  박혁  ;  박창서  ;  김기덕  ;  이희철 
Citation
 Korean Journal of Oral and Maxillofacial Radiology (대한구강악안면방사선학회지), Vol.33(3) : 171-178, 2003 
Journal Title
Korean Journal of Oral and Maxillofacial Radiology(대한구강악안면방사선학회지)
ISSN
 1229-8212 
Issue Date
2003
MeSH
tomography ; x-ray computed ; cephalometry ; reproducibility of results
Keywords
tomography ; x-ray computed ; cephalometry ; reproducibility of results
Abstract
PURPOSE : The purpose of this study was to report the reproducibility of intra-observer and inter-observer consistency of cephalometric measurements using three-dimensional (3D) computed tomography (CT), and the degree of difference of the cephalometric measurements.
MATERIALS AND METHODS : CT images of 16 adult patients with normal class I occlusion were sent to personal computer and reconstructed into 3D images using V-Works 3.5TM (Cybermed Inc., Seoul, Korea). With the internal program of V-Works 3.5TM, 12 landmarks on regular cephalograms were transformed into 21 analytic categories and measured by 2 observers and in addition, one of the observers repeated their measurements. Intra-observer difference was calculated using paired t-test, and inter-observer by two sample test.
RESULTS : There were significant differences in the intra-observer measurements (p<0.05) in four of the categories which included ANS-Me, ANS-PNS, Cdl-Go (Lt), GoL-GoR, but with the exception of Cdl-Go (Lt), ZmL-ZmR,
Zyo-Zyo, the average differences were within 2 mm of each other. The inter-observer observations also showed significant differences in the measurements of the ZmL-ZmR and Zyo-Zyo categories (p<0.05). With the exception of the Cdl-Me (Rt), ZmL-ZmR, Zyo-Zyo categories, the average differences between the two observers were within 2 mm, but the ZmL-ZmR and Zyo-Zyo values differed greatly with values of 8.10 and 19.8 mm respectively.
CONCLUSION : In general, 3D CT images showed greater accuracy and reproducibility, with the exception of suture areas such as Zm and Zyo, than regular cephalograms in orthodontic measurements, showing differences of less than 2 mm, therefore 3D CT images can be useful in cephalometric measurements and treatment planning.
Files in This Item:
T200305754.pdf Download
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Advanced General Dentistry (통합치의학과) > 1. Journal Papers
Yonsei Authors
Kim, Kee Deog(김기덕) ORCID logo https://orcid.org/0000-0003-3055-5130
Jeon, Kug Jin(전국진) ORCID logo https://orcid.org/0000-0002-5862-2975
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/114200
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.