490 772

Cited 0 times in

유방촬영술의 임상영상평가 : 전국조사

Other Titles
 Clinical Image Evaluation of Mammograms: A National Survey 
Authors
 문우경  ;  차주희  ;  임정기  ;  조나리야  ;  정선양  ;  김미혜  ;  이유진  ;  최은완  ;  조경수  ;  김태정  ;  연경모  ;  정수영  ;  최혜영  ;  김은경  ;  최연현  ;  한부경 
Citation
 Journal of the Korean Radiologist Society (대한방사선의학회지), Vol.49(6) : 507-511, 2003 
Journal Title
Journal of the Korean Radiologist Society(대한방사선의학회지)
ISSN
 0301-2867 
Issue Date
2003
Abstract
Purpose: The goal of this study was to survey the overall quality of mammographic images in Korea. Materials and Methods: A total of 598 mammographic images collected from 257 hospitals nationwide were reviewed in terms of eight image quality categories, namely positioning, compression, contrast, exposure, sharpness, noise, artifacts, and examination identification, and rated on a five-point scale: (1=severe deficiency, 2=major deficiency, 3=minor deficiency, 4=good, 5=best). Failure was defined as the occurrence of more than four major deficiencies or one severe deficiency (score of 1 or 2). The results were compared among hospitals of varying kinds, and common problems in clinical image quality were identified. Results: Two hundred and seventeen mammographic images (36.3%) failed the evaluation. Poor images were found in descending order of frequency, at The Society for Medical Examination (33/69, 47.8%), non-radiology clinics (42/88, 47.7%), general hospitals (92/216, 42.6%), radiology clinics (39/102, 38.2%), and university hospitals (11/123, 8.9%) (p<0.01, Chi-square test). Among the 598 images, serious problems which occurred were related to positioning in 23.7% of instances (n=142) (p<0.01, Chi-square test), examination identification in 5.7% (n=34), exposure in 5.4% (n=32), contrast in 4.2% (n=25), sharpness in 2.7% (n=16), compression in 2.5% (n=15), artifacts in 2.5% (n=15), and noise in 0.3% (n=2). Conclusion: This study showed that in Korea, 36.3% of the mammograms examined in this sampling had important image-related defects that might have led to serious errors in patient management. The failure rate was significantly higher in non-radiology clinics and at The Society for Medical Examination than at university hospitals
Files in This Item:
T200304905.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eun-Kyung(김은경) ORCID logo https://orcid.org/0000-0002-3368-5013
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/113958
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.