360 576

Cited 0 times in

새로이 개발된 경피적 삽입술이 가능한 대동맥 인조혈관 스텐트의 임상 시험

Other Titles
 Percutaneous Endovascular Stent-Graft Treatment of Aortic Aneurysms and Dissections: New Techniques and Initial Experience 
Authors
 이도연  ;  강성권  ;  원제환  ;  최동훈  ;  송호영  ;  원종윤  ;  이광훈 
Citation
 Journal of the Korean Radiologist Society (대한방사선의학회지), Vol.48(1) : 13-21, 2003 
Journal Title
Journal of the Korean Radiologist Society(대한방사선의학회지)
ISSN
 0301-2867 
Issue Date
2003
Abstract
Purpose: To evaluate the feasibility, safety and effectiveness of a newly designed percutaneously implanted separate stent-graft (SSG) for the treatment of aortic aneurysms and dissections.

Materials and Methods: Using a percutaneous technique, SSG placement (in the descending thoracic aorta in 26 cases and infrarenal abdominal aorta in 24) was attempted in 50 patients with aortic aneurysms (n=27) or dissection (n=23). All SSGs were individually constructed using self-expandable nitinol stents and a Dacron graft, and were introduced through a 12 F sheath and expanded to a diameter of 20-34 mm. In all cases, vascular access was through the femoral artery. The clinical status of each patient was monitored, and postoperative CT was performed within one week of the procedure and at 3-6 month intervals afterwards.

Results: Endovascular stent-graft deployment was technically successful in 49 of 50 patients (98%). The one failure was due to torsion of the unsupported graft during deployment. Successful exclusion of aneurysms and the primary entry tears of dissections was achieved in all but three patients with aortic dissection. All patients in whom technical success was achieved showed complete thrombosis of the thoracic false lumen or aneurysmal sac, and the overall technical success rate was 92%. In addition, sixteen patients demonstrated complete resolution of the dissected thoracic false lumen (n=9) or aneurysmal sac (n=7). Immediate post-operative complications occurred at the femoral puncture site in one patient with an arteriovenous fistula, and in two, a new saccular aneurysm developed at the distal margin of the stent. No patient died, and there was no instance of paraplegia, stroke, side-branch occlusion or infection during the subsequent mean follow-up period of 9.4 (range, 2 to 26) months.

Conclusion: In patients with aortic aneurysm and dissection, treatment with a separate percutaneously inserted stent-graft is technically feasible, safe, and effective.
Files in This Item:
T200302731.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
Yonsei Authors
Won, Jong Yun(원종윤) ORCID logo https://orcid.org/0000-0002-8237-5628
Lee, Kwang Hun(이광훈)
Lee, Do Yun(이도연)
Choi, Dong Hoon(최동훈) ORCID logo https://orcid.org/0000-0002-2009-9760
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/113370
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links