512 423

Cited 0 times in

Immunoassay분과 신빙도조사 결과보고 (2003)

Other Titles
 Annual Report on External Quality Assessment of Immunoassay Subcommittee in Korea (2003) 
Authors
 김현숙  ;  임환섭  ;  선영규  ;  김진규  ;  김영기  ;  김덕언  ;  권오헌  ;  정영순  ;  신섭  ;  전혜선  ;  조한수  ;  김영란 
Citation
 Journal of Laboratory Medicine and Quality Assurance (임상검사와 정도관리), Vol.26(1) : 103-121, 2004 
Journal Title
Journal of Laboratory Medicine and Quality Assurance(임상검사와 정도관리)
ISSN
 1225-097X 
Issue Date
2004
Keywords
Quality assurance ; External quality assessment ; Immunoassay
Abstract
Two trials of external quality assessment were performed in 2003. Thirteen test items of immunoassay with ten control materials were surveyed. The response rate of external quality assessment for Immunoassay Subcommittee were 93.8% and 92.8%. Ten control materials were consisted of 8 home-made pooled sera and 2 commercial control sera. The results are summarized as follows. 1. Laboratories participating in external quality control program of immunoassay were 259 laboratories and the response rate were 93.8% and 92.8% in 2003. 2. Chemiluminiscence immunoassay autoanalyzer was now widely introduced comparing to previous years and now it is the most popular analyzer in the field of immunoassay testing. 3. Still some test items show big variations of the test results of the same control material according to autoanalyzer. Generally the quality of the participating laboratories seems to be thought being improved. And in the following years, new planning of the statistic analysis and some standardization protocols could be introduced.
Files in This Item:
T200401052.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Hyon Suk(김현숙) ORCID logo https://orcid.org/0000-0001-5662-7740
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/112498
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links