Abstract
Purpose: We designed a water-based bolus device for radiation therapy in Kaposi's sarcoma. This study evaluated the usefulness of this new device and compared it with the currently used rice-based bolus. Materials and Methods: We fashioned a polystyrene box and cut a hole in order to insert patient's extremities while the patient was in the supine position. We used a vacuum-vinyl based polymer to reduce water leakage. Next, we eliminated air using a vacuum pump and a vacuum valve to reduce the air gap between the water and extremities in the vacuum-vinyl box. We performed CT scans to evaluate the density difference of the fabricated water-based bolus device when the device in which the rice-based bolus was placed directly, the rice-based bolus with polymer-vinyl packed rice, and the water were all put in. We analyzed the density change with the air gap volume using a planning system. In addition, we measured the homogeneity and dose in the low-extremities phantom, attached to six TLD, and wrapped film exposed in parallel-opposite fields with the LINAC under the same conditions as the set-up of the CT-simulator. Results: The density value of the rice-based bolus with the rice put in directly was 14% lower than that of the water-based bolus. Moreover, the value of the other experiments in the rice-based bolus with the polymer-vinyl packed rice showed an 18% reduction in density. The analysis of the EDR2 film revealed that the water-based bolus shows a more homogeneous dose plan, which was superior by $4{\sim}4.4%$ to the rice-base bolus. The mean TLD readings of the rice-based bolus, with the rice put directly into the polystyrene box had a 3.4% higher density value. Moreover, the density value in the case of the rice-based bolus with polymer-vinyl packed rice had a 4.3% higher reading compared to the water-based bolus. Conclusion: Our custom-made water-based bolus device increases the accuracy of the set-up by confirming the treatment field. It also improves the accuracy of the therapy owing to the reduction of the air gap using a vacuum pump and a vacuum valve. This set-up represents a promising alternative device for delivering a homogenous dose to the target volume
Files in This Item:
T200801043.pdf Download
DOI
10.3857/jkstro.2008.26.3.189
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, Yong Bae(김용배) ORCID logo https://orcid.org/0000-0001-7573-6862
Suh, Chang Ok(서창옥)
Lee, Ik Jae(이익재) ORCID logo https://orcid.org/0000-0001-7165-3373
Cho, Jae Ho(조재호) ORCID logo https://orcid.org/0000-0001-9966-5157
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/107314
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links