712 938

Cited 0 times in

웨이브프론트 방식을 이용한 라섹 수술 시 수술 전 다양한 굴절 검사의 정확도 비교

Other Titles
 Comparison of the Refractive Error Measurement Using Different Methods in Wavefront-Guided LASEK 
Authors
 안지민  ;  안현석  ;  서경률  ;  김응권  ;  김태임 
Citation
 JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, Vol.51(9) : 1196-1202, 2010 
Journal Title
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY(대한안과학회지)
ISSN
 0378-6471 
Issue Date
2010
Keywords
iTrace aberrometer ; Nomogram ; Postcycloplegic refraction ; Wavefront-guided LASEK ; Wavescan
Abstract
PURPOSE: To predict the accuracy of preoperative refractive error measurement methods in wavefront-guided laser-assisted subepithelial keratectomy (LASEK) surgery and to formulate a nomogram for satisfactory surgical results. METHODS: The medical records of 30 patients (57 eyes) who had undergone wavefront-guided LASEK were reviewed. The ideal surgical ablation amount was defined as the sum of the real surgical ablation amount and the remaining refractive errors. Comparison between the ideal surgical ablation amount and preoperative refractive errors was made using autorefraction, manifest refraction, cycloplegic refraction, postcycloplegic refraction, wavescan, and iTrace aberrometer measurements. RESULTS: The refractive errors measured by the postcycloplegic refraction showed the closest relation with the ideal surgical amount, and the nomogram based on this refraction correlated statistically significantly with the ideal surgical ablation amount. The refractive error using the wavescan also showed more accurate refractive measurements than the autorefractor and iTrace aberrometer. CONCLUSIONS: Accurate manifest refraction immediately before surgery is the most important in determining the ablation amount. Additionally, the refractive errors measured with the wavescan, which is an aberrometer used for wavefront-guided LASEK, showed a minimal amount of errors. After reviewing the results, the nomogram based on these 2 methods can be concluded to possibly contribute to an increase in the accuracy of surgery
Files in This Item:
T201003383.pdf Download
DOI
10.3341/jkos.2010.51.9.1196
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Ahn, Hyun Seok(안현석)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/102141