519 731

Cited 0 times in

기관편향이 심한 환자에서 두부 위치 변경 후 발생한 기관내관의 폐쇄

Other Titles
 Obstruction of an Endotracheal Tube with Change of the Head Position in a Trachea Deviated Patient 
Authors
 한동우  ;  장철호  ;  남상범  ;  이수현  ;  나성원  ;  이종석 
Citation
 KOREAN JOURNAL OF ANESTHESIOLOGY, Vol.52(3) : 339-341, 2007 
Journal Title
KOREAN JOURNAL OF ANESTHESIOLOGY(대한마취과학회지)
ISSN
 2005-6419 
Issue Date
2007
Abstract
Various causes of endotracheal tube obstruction during general anesthesia exist. Herein, the case of a 77-year old male patient, with tracheal deviation, who was intubated with an endotracheal tube in the emergency room, without a Murphy eye, is reported. He was transferred to the operating room for the removal of a subdural hematoma. The endotracheal tube suddenly became obstructed during flexion of neck for the craniotomy procedure. The distal bevel of the endotracheal tube, without Murphy eye, was discovered to have come into contact with the tracheal wall, with the airway being obstructed during the fiberoptic bronchoscopy. Shortly after extubation of the obstructed tube, an armored tube, with a Murphy eye, was reintubated, after which the respiration pattern immediately returned to normal.
Files in This Item:
T200700190.pdf Download
DOI
10.4097/kjae.2007.52.3.339
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers
Yonsei Authors
Na, Sungwon(나성원) ORCID logo https://orcid.org/0000-0002-1170-8042
Nam, Sang Beom(남상범) ORCID logo https://orcid.org/0000-0002-9704-1866
Lee, Jong Seok(이종석) ORCID logo https://orcid.org/0000-0002-7945-2530
Chang, Chul Ho(장철호) ORCID logo https://orcid.org/0000-0001-5647-8298
Han, Dong Woo(한동우) ORCID logo https://orcid.org/0000-0002-8757-663X
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/96812
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links