A-Se 기반 디지털 X-선 영상장치의 Contrast-detail 특성 평가

Other Titles
 Evaluation of Contrast-detail Characteristics of an A-Se Based Digital X-ray Imaging System 
Authors
 현혜경  ;  박소현  ;  조효성  ;  조희문  ;  김근영 
Citation
 Journal of the Korean Society of Radiology (한국방사선학회 논문지), Vol.1(1) : 11-16, 2007 
Journal Title
Journal of the Korean Society of Radiology(한국방사선학회 논문지)
ISSN
 1976-0620 
Issue Date
2007
Keywords
A-Se ; Digital X-ray ; Contrast-detail
Abstract
In this study, we have performed contrast-detail analysis for an amorphous selenium(a-Se) based digital X-ray imaging system by using a contrast-detail phantom(CDRAD 2.0) to test its low contrast performance. The X-ray imaging system utilizes an 500-mm-thick a-Se semiconductor X-ray absorber coated over an amorphous silicon(a-Si) TFT(thin-film transistor) detector matrix with a $139mm{\times}139mm$ pixel size and a $46.7cm{\times}46.7cm$ active area. In the measurement of contrast-detail curves we first acquired X-ray images of the CDRAD 2.0 phantom at given test conditions(i.e., 40, 50, 60, 70, 80 kVp, and 16 mA.s), and then evaluated the contrast-detail characteristics of the imaging system from each phantom image by using an image quality factor called the image-quality-figure-inverse(IQFinv). The IQFinv values for the imaging system gradually improved with the photon fluence, indicating the improvement of image visibility: 24.4, 35.3, 39.2, 41.5, and 43.4 at photon fluences of $1.8{\times}105$, $5.9{\times}105$, $11.3{\times}105$, $19.4{\times}105$, and $29.4{\times}105$ photons/$mm^2$, respectively.
Files in This Item:
T200704688.pdf Download
Appears in Collections:
3. College of Nursing (간호대학) > Dept. of Nursing (간호학과) > 1. Journal Papers
Yonsei Authors
Park, So Hyun(박소현)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/96689
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links