302 509

Cited 2 times in

Inter-Device Agreement of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Cirrus HD OCT

Authors
 Samin Hong  ;  Yeongjun Kim  ;  ongwoo Shim  ;  Chan Yun Kim  ;  Gong Je Seong 
Citation
 Korean Journal of Ophthalmology, Vol.25(2) : 105-109, 2011 
Journal Title
Korean Journal of Ophthalmology
ISSN
 1011-8942 
Issue Date
2011
MeSH
Inter-deviceagreement ; Optical coherence tomography ; Retinalnervefiberlayerthickness
Keywords
Adult ; Algorithms* ; Female ; Humans ; Male ; Observer Variation ; OpticNerveDiseases/diagnosis* ; RetinalGanglion Cells/pathology* ; Tomography, Optical Coherence/methods ; Tomography, Optical Coherence/statistics & numerical data* ; Young Adult
Abstract
PURPOSE: To assess the inter-device agreement of peripapillary retinal nerve fiber layer (RNFL) thickness measurements by 2 spectral domain Cirrus HD optical coherence tomography (OCT) devices in healthy Korean subjects. METHODS: Eleven eyes of 11 healthy volunteers were enrolled in the present study. Each eye was scanned with the Optic Disc Cube 200 x 200 scan of 2 Cirrus HD OCT devices for peripapillary RNFL thickness calculation. The inter-device agreements of the 2 Cirrus HD OCTs for average, quadrant, and clock-hour RNFL thickness values were determined with Wilcoxon signed rank test, Friedman test, Cronbach's alpha (alpha), intraclass correlation coefficient (ICC), coefficient of variation (COV), and Bland-Altman plot. RESULTS: The mean age of the participants was 25.82 +/- 3.28 years and all had a 0.00 logarithm of the minimum angle of resolution of best-corrected visual acuity. The signal strengths of scans from the 2 Cirrus HD OCT were not significantly different (p = 0.317). The inter-device agreement of average RNFL thickness was excellent (alpha, 0.940; ICC, 0.945; COV, 2.45 +/- 1.52%). However, the agreement of nasal quadrant RNFL thickness was not very good (alpha, 0.715; ICC, 0.716; COV, 5.72 +/- 4.64%). Additionally, on the Bland-Atman plot, the extent of agreement of the 2 Cirrus HD OCTs for RNFL thickness was variable according to scanned sectors. CONCLUSIONS: The inter-device agreement of 2 spectral domain Cirrus HD OCT devices for peripapillary RNFL thickness measurements was generally excellent but variable according to the scanned area. Thus, physicians should consider this fact before judging a change of RNFL thicknesses if they were measured by different OCT devices.
Files in This Item:
T201103296.pdf Download
DOI
10.3341/kjo.2011.25.2.105
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Chan Yun(김찬윤) ORCID logo https://orcid.org/0000-0002-8373-9999
Seong, Gong Je(성공제) ORCID logo https://orcid.org/0000-0002-5456-4296
Hong, Sa Min(홍사민)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/94152
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links