5 6

Cited 0 times in

Cited 0 times in

Comparative evaluation of two NGS-based assays for somatic hypermutation analysis of IGHV genes in chronic lymphocytic leukemia

Authors
 Kang, Yehyun  ;  Lee, Hyeonah  ;  Park, Yu Jin  ;  Lee, Seung-Tae  ;  Choi, Jong Rak  ;  Shin, Saeam 
Citation
 BLOOD RESEARCH, Vol.61(1), 2026-04 
Article Number
 14 
Journal Title
BLOOD RESEARCH
ISSN
 2287-979X 
Issue Date
2026-04
Keywords
Chronic lymphocytic leukemia ; Somatic hypermutation ; Sanger sequencing ; Next-generation sequencing ; Immunoglobulin heavy chain gene
Abstract
PurposeSomatic hypermutation (SHM) of the immunoglobulin heavy chain variable (IGHV) region is a key prognostic marker in chronic lymphocytic leukemia (CLL). Traditionally, SHM status has been determined using Sanger sequencing (SS); however, next-generation sequencing (NGS) provides an alternative method for assessing both SHM status and clonal rearrangements. This study aimed to compare the performance of two commercially available NGS assays for evaluating IGH clonality and SHM status in CLL.MethodsIn this retrospective study, 42 samples from patients diagnosed with CLL were analyzed. Genomic DNA extracted from peripheral blood or bone marrow aspirates was analyzed using two commercial NGS assays: the LymphoTrack (R) IGHV Leader (Leader) and IGH FR1 (FR1) assays (Invivoscribe, CA, USA). SS was performed as the reference method for SHM assessment.ResultsThe Leader assay identified clonality in 95.2% of cases, whereas the FR1 assay detected clonality in 88.1%. Conclusive SHM status was determined in 90.5% of samples using the Leader assay and in 76.2% using the FR1 assay; when results from both assays were combined, the rate increased to 92.9%. Among samples with conclusive results by both SS and each NGS assay, the Leader assay demonstrated higher concordance with SS (97.1%, 34/35) than the FR1 assay (86.2%, 25/29). Greater variability in clonal detection was observed with the FR1 assay.ConclusionThese findings indicate that the Leader assay provides a more reliable assessment of SHM status, with higher concordance with SS. Although the FR1 assay may offer additional information regarding clonal patterns, its results should be interpreted cautiously. Given the limited sample size, further studies are warranted to validate these findings. Overall, the Leader assay appears to be more suitable as a primary tool for SHM evaluation, with FR1 results serving a complementary role when interpreted in clinical context.
Files in This Item:
92488.pdf Download
DOI
10.1007/s44313-026-00132-7
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Park, Yu Jin(박유진)
Shin, Saeam(신새암) ORCID logo https://orcid.org/0000-0003-1501-3923
Lee, Seung-Tae(이승태) ORCID logo https://orcid.org/0000-0003-1047-1415
Choi, Jong Rak(최종락) ORCID logo https://orcid.org/0000-0002-0608-2989
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/212030
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links