Cited 0 times in

Validation of Devices for the Five Times Sit To Stand Test: Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement

Authors
 Sanghyun Jee  ;  Chan Woong Jang  ;  Kyoungmin Park  ;  Sanghoon Shin  ;  Min-Chul Paek  ;  Jung Hyun Park 
Citation
 YONSEI MEDICAL JOURNAL, Vol.66(1) : 51-57, 2025-01 
Journal Title
YONSEI MEDICAL JOURNAL
ISSN
 0513-5796 
Issue Date
2025-01
MeSH
Aged ; Aged, 80 and over ; Female ; Foot* / physiology ; Head / physiology ; Humans ; Male ; Middle Aged ; Pressure* ; Standing Position
Keywords
Aged ; foot orthoses ; frailty ; motion capture ; physical functional performance
Abstract
Purpose: This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.

Materials and methods: Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.

Results: Participants' mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).

Conclusion: This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
Files in This Item:
T992025031.pdf Download
DOI
10.3349/ymj.2023.0525
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Rehabilitation Medicine (재활의학교실) > 1. Journal Papers
Yonsei Authors
Park, Jung Hyun(박중현) ORCID logo https://orcid.org/0000-0003-3262-7476