Image quality-based dose optimization in pediatric cone-beam computed tomography: A pilot methodological study

Authors
 Hak-Sun Kim  ;  Yoon Joo Choi  ;  Kug Jin Jeon  ;  Sang-Sun Han  ;  Chena Lee 
Citation
 IMAGING SCIENCE IN DENTISTRY, Vol.54(3) : 264-270, 2024-09 
Journal Title
IMAGING SCIENCE IN DENTISTRY
ISSN
 2233-7822 
Issue Date
2024-09
Keywords
Cone-Beam Computed Tomography ; Dose Reduction ; Radiation Dosage ; Radiation Protection
Abstract
Purpose: This study aimed to propose a methodological approach for reducing the radiation dose in pediatric conebeam computed tomography (CBCT), focusing exclusively on balancing image quality with dose optimization.

Materials and methods: The dose-area product (DAP) for exposure was reduced using copper-plate attenuation of an X-ray source. The thickness of copper (Cu) was increased from 0 to 2.2 mm, and 10 different DAP levels were used. The QUART DVT_AP phantom and pediatric radiologic dentiform were scanned under the respective DAP levels. The contrast-to-noise ratio (CNR), image homogeneity, and modulation transfer function (MTF) were analyzed using the QUART DVT_AP phantom. An expert evaluation (overall image grade, appropriateness of field of view, artifacts, noise, and resolution) was conducted using pediatric dentiform images. The critical DAP level was determined based on phantom and dentiform analysis results.

Results: CNR and image homogeneity decreased as the DAP was reduced; however, there was an inflection point of image homogeneity at Cu 1.6 mm (DAP=138.00 mGy·cm2), where the value started increasing. The MTF showed constant values as the DAP decreased. The expert evaluation of overall image grades showed "no diagnostic value" for dentiform images with Cu 1.9-2.2 mm (DAP=78.00-103.33 mGy·cm2). The images with Cu 0-1.6 mm (DAP=138.00-1697.67 mGy·cm2) had a "good," "moderate," or "poor but interpretable" grade.

Conclusion: Reducing DAP beyond a 1.6-mm Cu thickness degraded CBCT image quality. Image homogeneity and clinical image grades indicated crucial decision points for DAP reduction in pediatric CBCT scans.
Files in This Item:
T992024636.pdf Download
DOI
10.5624/isd.20240030
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Oral and Maxillofacial Radiology (영상치의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Hak-Sun(김학선) ORCID logo https://orcid.org/0000-0002-6833-6577
Lee, Chena(이채나) ORCID logo https://orcid.org/0000-0002-8943-4192
Jeon, Kug Jin(전국진) ORCID logo https://orcid.org/0000-0002-5862-2975
Choi, Yoon Joo(최윤주) ORCID logo https://orcid.org/0000-0001-9225-3889
Han, Sang Sun(한상선) ORCID logo https://orcid.org/0000-0003-1775-7862
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/201758
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links