Cited 0 times in

Development of a new VMAT QA framework for Mobius3D using control-point specific EPID images

Authors
 JaeHyun Seok  ;  Hojin Kim  ;  Min Cheol Han  ;  Jihun Kim  ;  Kwangwoo Park  ;  Hyeonjeong Cho  ;  Dohyeon Yoo  ;  Jin Sung Kim 
Citation
 FRONTIERS IN ONCOLOGY, Vol.14 : 1478118, 2024-12 
Journal Title
FRONTIERS IN ONCOLOGY
Issue Date
2024-12
Keywords
EPID ; VMAT QA ; control-point specific QA ; log-based QA ; patient-specific QA
Abstract
Purpose: This study presents novel quality assurance (QA) approach for volumetric modulated arc therapy (VMAT) that leverages frame-by-frame electronic portal imaging device (EPID) images integrated into Mobius3D for accurate three-dimensional dose calculations.

Methods: Sequential EPID images for VMAT plans were acquired every 0.4-second by iView system and processed through iterative deconvolution to mitigate blurring from photon scattering. Deconvolved images were binarized to define multi-leaf collimator (MLC) positions. Pre-acquired box fluences determined optimal threshold for binarization and adjusted for detector shift depending on gantry and collimator angles. Sequential EPID images were re-scaled using pixel scaling factor (PSF) and converted to monitor unit (MU) proportional values. Generated EPID-based log file, including control-point specific MLC and monitor units (MU) information, were analyzed in Mobius3D for Gamma passing rate (GPR) of VMAT plans from 18 patients. Plan complexity indices were calculated and correlated with GPR.

Results: Clinically appropriate threshold was defined to be 20000 that can extract accurate MLC data from the deconvolved binarized EPID images. Positional deviations due to gantry and collimator rotations were observed to be up to 4.5 pixels. Recalibrated EPID pixel values showed linearity with MU regardless of changes in dose rate. Consequently, average GPR for 18 patients evaluated using Mobius3D reached 95.2% ± 3.7%%, based on 3% dose difference and 3mm distance-to-agreement criterion. It was found that two plan complexity indices showed statistically significant correlation with GPR.

Conclusion: This study successfully implemented novel measurement-based VMAT QA framework based on control-point specific EPID, based upon accurate MLC and MU data at each frame.
Files in This Item:
T202407484.pdf Download
DOI
10.3389/fonc.2024.1478118
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jihun(김지훈) ORCID logo https://orcid.org/0000-0003-4856-6305
Kim, Jinsung(김진성) ORCID logo https://orcid.org/0000-0003-1415-6471
Kim, Hojin(김호진) ORCID logo https://orcid.org/0000-0002-4652-8682
Park, Kwang Woo(박광우) ORCID logo https://orcid.org/0000-0002-9843-7985
Yoo, Dohyeon(유도현)
Han, Min Cheol(한민철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/201541
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links