Cited 0 times in

Clinical validity and precision of deep learning-based cone-beam computed tomography automatic landmarking algorithm

Authors
 Jungeun Park  ;  Seongwon Yoon  ;  Hannah Kim  ;  Youngjun Kim  ;  Uilyong Lee  ;  Hyungseog Yu 
Citation
 IMAGING SCIENCE IN DENTISTRY, Vol.54(3) : 240-250, 2024-08 
Journal Title
IMAGING SCIENCE IN DENTISTRY
ISSN
 2233-7822 
Issue Date
2024-08
Keywords
Anatomic Landmarks ; Cephalometry ; Cone-Beam Computed Tomography ; Deep Learning ; Orthognathic Surgery
Abstract
Purpose: This study was performed to assess the clinical validity and accuracy of a deep learning-based automatic landmarking algorithm for cone-beam computed tomography (CBCT). Three-dimensional (3D) CBCT head measurements obtained through manual and automatic landmarking were compared.

Materials and Methods: A total of 80 CBCT scans were divided into 3 groups: non-surgical (39 cases); surgical without hardware, namely surgical plates and mini-screws (9 cases); and surgical with hardware (32 cases). Each CBCT scan was analyzed to obtain 53 measurements, comprising 27 lengths, 21 angles, and 5 ratios, which were determined based on 65 landmarks identified using either a manual or a 3D automatic landmark detection method. Results: In comparing measurement values derived from manual and artificial intelligence landmarking, 6 items displayed significant differences: R U6CP-L U6CP, R L3CP-L L3CP, S-N, Or_R-R U3CP, L1L to Me-GoL, and GoR-Gn/S-N (P<0.05). Of the 3 groups, the surgical scans without hardware exhibited the lowest error, reflecting the smallest difference in measurements between human- and artificial intelligence-based landmarking. The time required to identify 65 landmarks was approximately 40-60 minutes per CBCT volume when done manually, compared to 10.9 seconds for the artificial intelligence method (PC specifications: GeForce 2080Ti, 64GB RAM, and an Intel i7 CPU at 3.6 GHz).

Conclusion: Measurements obtained with a deep learning-based CBCT automatic landmarking algorithm were similar in accuracy to values derived from manually determined points. By decreasing the time required to calculate these measurements, the efficiency of diagnosis and treatment may be improved.
Files in This Item:
T202405666.pdf Download
DOI
10.5624/isd.20240009
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Orthodontics (교정과학교실) > 1. Journal Papers