0 13

Cited 0 times in

Correction method for depth and field size dependence in Octavius 1000 SRS patient-specific QA

Authors
 Han-Back Shin  ;  Yu-Yun Noh  ;  Heesoo Kim  ;  Changhwan Kim  ;  KiHoon Sung  ;  Dong Wook Kim 
Citation
 RADIATION PHYSICS AND CHEMISTRY, Vol.223 : 111892, 2024-06 
Journal Title
RADIATION PHYSICS AND CHEMISTRY
ISSN
 0969-806X 
Issue Date
2024-06
Keywords
Octavius 1000 SRS ; Patient-specific quality assurance ; Depth dependent ; Small field size dosimetry ; QA correction method
Abstract
Stereotactic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) are advanced techniques capable of delivering higher doses of radiation precisely to small and complex targets. As these higher doses are delivered, the significance of patient-specific quality assurance becomes increasingly crucial to ensure both the effectiveness and safety of treatment. This study investigates the depth and field size dependence in Octavius 1000 SRS patient-specific quality assurance (PSQA), which is vital for ensuring accuracy in modern radiotherapy treatments. It comprehensively compares the performance of the Octavius 1000 SRS array against standard treatment planning system calculations, focusing on various field sizes and depths. An extensive measurement process was performed, wherein output factors were analyzed across a range of field sizes, from very small (1 × 1 cm2) to larger fields (10 × 10 cm2), at different depths (5 cm, 10 cm, and 15 cm). Our results highlighted significant variations in the measurement accuracy, dependent on the field size and depth, which underscored the challenge involved in ensuring precise dosimetry in PSQA for small fields. The largest discrepancy observed was 2.65% for a 1 × 1 cm2 field size at a depth of 15 cm. Based on these findings, a novel correction method tailored for small field sizes and varying depths was proposed. This method enhances the accuracy and reliability of PSQA in advanced radiotherapy to ensure that the delivered dose conforms closely to the planned dose, which is a critical aspect in treatments targeting small and complex tumor volumes. Through the proposed correction, this study aims to bridge the gap between calculated and measured dose distributions, thus contributing significantly to the refinement of PSQA processes in radiotherapy. © 2024 The Authors
Full Text
https://www.sciencedirect.com/science/article/pii/S0969806X24003840
DOI
10.1016/j.radphyschem.2024.111892
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers