Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 이미정 | - |
dc.date.accessioned | 2023-08-09T02:35:50Z | - |
dc.date.available | 2023-08-09T02:35:50Z | - |
dc.date.issued | 2017-04 | - |
dc.identifier.issn | 0899-7071 | - |
dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/195663 | - |
dc.description.abstract | This study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64±0.15 μGy and 49.11±1.46 μGy, respectively (p<0.001). The mean SNR values for mini-DI and DR were 233.2±5.1 and 31.6±1.2, and 10% MTF values were 131 and 161μm. A newly developed mini-DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care. | - |
dc.description.statementOfResponsibility | restriction | - |
dc.language | English | - |
dc.publisher | Elsevier | - |
dc.relation.isPartOf | CLINICAL IMAGING | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.subject.MESH | Humans | - |
dc.subject.MESH | Infant, Newborn | - |
dc.subject.MESH | Intensive Care Units, Neonatal* | - |
dc.subject.MESH | Phantoms, Imaging | - |
dc.subject.MESH | Radiation Dosage | - |
dc.subject.MESH | Radiographic Image Enhancement / methods* | - |
dc.subject.MESH | Radiography, Thoracic / methods* | - |
dc.subject.MESH | Thorax / diagnostic imaging* | - |
dc.title | Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit | - |
dc.type | Article | - |
dc.contributor.college | College of Medicine (의과대학) | - |
dc.contributor.department | Dept. of Radiology (영상의학교실) | - |
dc.contributor.googleauthor | Tae-Hoon Kim | - |
dc.contributor.googleauthor | Jong-Hyun Ryu | - |
dc.contributor.googleauthor | Chang-Won Jeong | - |
dc.contributor.googleauthor | Hong Young Jun | - |
dc.contributor.googleauthor | Dong-Woon Heo | - |
dc.contributor.googleauthor | Seung Hyun Lee | - |
dc.contributor.googleauthor | Yeon Kyun Oh | - |
dc.contributor.googleauthor | Mi-Jung Lee | - |
dc.contributor.googleauthor | Kwon-Ha Yoon | - |
dc.identifier.doi | 10.1016/j.clinimag.2016.12.004 | - |
dc.contributor.localId | A02774 | - |
dc.relation.journalcode | J00577 | - |
dc.identifier.eissn | 1873-4499 | - |
dc.identifier.pmid | 28064140 | - |
dc.identifier.url | https://www.sciencedirect.com/science/article/pii/S089970711630198X | - |
dc.subject.keyword | Chest X-ray imaging | - |
dc.subject.keyword | Conventional mobile digital radiography (DR) | - |
dc.subject.keyword | Mini-mobile digital imaging system (mini-DI) | - |
dc.subject.keyword | Premature neonate | - |
dc.contributor.alternativeName | Lee, Mi Jung | - |
dc.contributor.affiliatedAuthor | 이미정 | - |
dc.citation.volume | 42 | - |
dc.citation.startPage | 165 | - |
dc.citation.endPage | 171 | - |
dc.identifier.bibliographicCitation | CLINICAL IMAGING, Vol.42 : 165-171, 2017-04 | - |