Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit

DC Field Value Language
dc.contributor.author이미정-
dc.date.accessioned2023-08-09T02:35:50Z-
dc.date.available2023-08-09T02:35:50Z-
dc.date.issued2017-04-
dc.identifier.issn0899-7071-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/195663-
dc.description.abstractThis study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64±0.15 μGy and 49.11±1.46 μGy, respectively (p<0.001). The mean SNR values for mini-DI and DR were 233.2±5.1 and 31.6±1.2, and 10% MTF values were 131 and 161μm. A newly developed mini-DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care.-
dc.description.statementOfResponsibilityrestriction-
dc.languageEnglish-
dc.publisherElsevier-
dc.relation.isPartOfCLINICAL IMAGING-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHHumans-
dc.subject.MESHInfant, Newborn-
dc.subject.MESHIntensive Care Units, Neonatal*-
dc.subject.MESHPhantoms, Imaging-
dc.subject.MESHRadiation Dosage-
dc.subject.MESHRadiographic Image Enhancement / methods*-
dc.subject.MESHRadiography, Thoracic / methods*-
dc.subject.MESHThorax / diagnostic imaging*-
dc.titleReduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiology (영상의학교실)-
dc.contributor.googleauthorTae-Hoon Kim-
dc.contributor.googleauthorJong-Hyun Ryu-
dc.contributor.googleauthorChang-Won Jeong-
dc.contributor.googleauthorHong Young Jun-
dc.contributor.googleauthorDong-Woon Heo-
dc.contributor.googleauthorSeung Hyun Lee-
dc.contributor.googleauthorYeon Kyun Oh-
dc.contributor.googleauthorMi-Jung Lee-
dc.contributor.googleauthorKwon-Ha Yoon-
dc.identifier.doi10.1016/j.clinimag.2016.12.004-
dc.contributor.localIdA02774-
dc.relation.journalcodeJ00577-
dc.identifier.eissn1873-4499-
dc.identifier.pmid28064140-
dc.identifier.urlhttps://www.sciencedirect.com/science/article/pii/S089970711630198X-
dc.subject.keywordChest X-ray imaging-
dc.subject.keywordConventional mobile digital radiography (DR)-
dc.subject.keywordMini-mobile digital imaging system (mini-DI)-
dc.subject.keywordPremature neonate-
dc.contributor.alternativeNameLee, Mi Jung-
dc.contributor.affiliatedAuthor이미정-
dc.citation.volume42-
dc.citation.startPage165-
dc.citation.endPage171-
dc.identifier.bibliographicCitationCLINICAL IMAGING, Vol.42 : 165-171, 2017-04-