81 187

Cited 0 times in

응급맥박산소측정기

Other Titles
 PULSE OXIMETRY IN THE EMERGENCY DEPARTMENT 
Authors
 조광현  ;  손대곤  ;  이한식 
Citation
 Journal of the Korean Society of Emergency Medicine (대한응급의학회지), Vol.5(1) : 87-92, 1994-07 
Journal Title
Journal of the Korean Society of Emergency Medicine(대한응급의학회지)
ISSN
 1226-4334 
Issue Date
1994-07
Abstract
Study objective : To determine the accuracy of the pulse oximetry in the emergency department. Study design:. Prospective, cross sectional, paired measurement of SpOz against SaOz. Setting: University affiliated hospital. Interventions. A pulse oximeter with digital probe was used to measure Spo2. Arterial blood gas was drawn at the same time to measure SaOz. Results. There were no significant difference between SpOz and SaOz when SpOz was above or less than 90%(mean>_90% =0.598+-2.665,mean<90% =2.926+-7.807;p>_90% =0. 0796,p<90% =0.0860), but all together there were significant difference(mean=1.22+_4.70; p = 0.017). Conclusion. The pulse oximetry is potentially useful in patient,s with clinical signs of acute hypoxemia and patients receiving interventions that may produce acute hypoxemia, and is a rnust device in emergency department.
Files in This Item:
T199401248.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Emergency Medicine (응급의학교실) > 1. Journal Papers
Yonsei Authors
Lee, Hahn Shick(이한식)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/195099
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links