Effect of Outlier Elimination on the 99th Percentile Upper Reference Limits of High-Sensitivity Cardiac Troponin I Assays Based on a Strictly Selected Healthy Reference Population

DC Field Value Language
dc.date.accessioned2023-06-02T00:50:25Z-
dc.date.available2023-06-02T00:50:25Z-
dc.date.issued2022-05-
dc.identifier.issn2234-3806-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/194466-
dc.description.abstractBackground: We established high-sensitivity cardiac troponin I (hsTnI) 99th percentile upper reference limits (URLs) for the Centaur XPT High-Sensitivity Troponin I assay (Centaur hsTnI; Siemens, Erlangen, Germany) and Atellica IM High-Sensitivity Troponin I assay (Atellica hsTnI; Siemens) and assessed the effect of outlier elimination. Methods: The reference population comprised 380 men and 387 women, satisfying the strict systematic reference population criteria. After reference population verification by the N-terminal pro-B-type natriuretic peptide (NT-proBNP) assay, 99th percentile URLs for Centaur hsTnI and Atellica hsTnI were calculated before and after outlier elimination. Results: The 99th percentile URL for Centaur hsTnI was 60.4 (men, 74.7; women, 57.5) ng/L and that for Atellica hsTnI was 59.6 (men, 75.2; women, 55.1) ng/L. After the elimination of 61 (8.0%) outlier samples in Centaur hsTnI and 58 (7.6%) in Atellica hsTnI, the 99th percentile URLs were 13.5 ng/L (men, 15.3 ng/L; women, 11.9 ng/L) and 13.4 ng/L (men, 15.5 ng/L; women, 12.9 ng/L), respectively, significantly lower than those before outlier elimination. The CVs at the 99th percentile URLs were 5.2% and 3.5%, respectively. The measurable fractions among the reference population were 91.5% and 93.4%, respectively. Performance evaluation of Atellica B-type natriuretic peptide (BNP), Atellica NT-proBNP, Centaur hsTnI, and Atellica hsTnI showed outstanding results. Conclusions: The Korean hsTnI 99th percentile URLs calculated in this study were significantly lower after outlier elimination than before. Centaur hsTnI and Atellica hsTnI meet the "Guideline acceptable" and "Level 3 (second generation, high sensitivity)" requirements, satisfying international standards.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherKorean Society for Laboratory Medicine-
dc.relation.isPartOfANNALS OF LABORATORY MEDICINE-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHAsian People*-
dc.subject.MESHBiomarkers-
dc.subject.MESHFemale-
dc.subject.MESHHumans-
dc.subject.MESHMale-
dc.subject.MESHReference Values-
dc.subject.MESHTroponin I* / analysis-
dc.titleEffect of Outlier Elimination on the 99th Percentile Upper Reference Limits of High-Sensitivity Cardiac Troponin I Assays Based on a Strictly Selected Healthy Reference Population-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Laboratory Medicine (진단검사의학교실)-
dc.contributor.googleauthorSunyoung Ahn-
dc.contributor.googleauthorHyun-Ki Kim-
dc.contributor.googleauthorWoochang Lee-
dc.contributor.googleauthorSail Chun-
dc.contributor.googleauthorWon-Ki Min-
dc.identifier.doi10.3343/alm.2022.42.3.331-
dc.relation.journalcodeJ00164-
dc.identifier.eissn2234-3814-
dc.identifier.pmid34907103-
dc.subject.keywordADVIA Centaur XPT High-Sensitivity Troponin I assay-
dc.subject.keywordAtellica IM High-Sensitivity Troponin I assay-
dc.subject.keywordKorean-
dc.subject.keywordNinety-ninth percentile-
dc.subject.keywordOutlier-
dc.subject.keywordUpper reference limit-
dc.citation.volume42-
dc.citation.number3-
dc.citation.startPage331-
dc.citation.endPage341-
dc.identifier.bibliographicCitationANNALS OF LABORATORY MEDICINE, Vol.42(3) : 331-341, 2022-05-