156 253

Cited 0 times in

Validity and Reliability of a Non-Radiographic Postural Analysis Device Based on an RGB-Depth Camera Comparing EOS 3D Imaging: A Prospective Observational Study

Authors
 Hyo Jeong Lee  ;  Han Eol Cho  ;  Myungsang Kim  ;  Seok Young Chung  ;  Jung Hyun Park 
Citation
 HEALTHCARE, Vol.11(5) : 686, 2023-02 
Journal Title
HEALTHCARE
Issue Date
2023-02
Keywords
full-body X-ray ; musculoskeletal pain ; posture ; skeleton ; three-dimensional imaging
Abstract
The posture-analyzing and virtual reconstructing device (PAViR) used a Red Green Blue-Depth camera as a sensor and skeleton reconstruction images were produced. This PAViR quickly analyzed the whole posture from multiple repetitive shots without radiation exposure in clothes and provided a virtual skeleton within seconds. This study aims to evaluate the reliability when shooting repeatedly and to assess the validity compared to parameters of full-body, low-dose X-rays (EOSs) when applied as diagnostic imaging. As a prospective and observational study, 100 patients with musculoskeletal pain underwent an EOS to obtain whole body coronal and sagittal images. The outcome measures were human posture parameters, which were divided by the standing plane in both EOSs and PAViRs as follows: (1) a coronal view (asymmetric clavicle height, pelvic oblique, bilateral Q angles of the knee, and center of seventh cervical vertebra-central sacral line (C7-CSL)) and (2) a sagittal view (forward head posture). A validation of the PAViR compared to the EOSs revealed that C7-CSL showed a moderate positive correlation with that of the EOS (r = 0.42, p < 0.01). The forward head posture (r = 0.39, p < 0.01), asymmetric clavicle height (r = 0.37, p < 0.01), and pelvic oblique (r = 0.32, p < 0.01) compared to those of the EOS had slightly positive correlations. The PAViR has excellent intra-rater reliability in people with somatic dysfunction. Except for both Q angles, the PAViR has fair-to-moderate validation when compared to EOS diagnostic imaging in the parameter representing coronal and sagittal imbalance. Although the PAViR system is not yet available in the medical field, it has the potential to become a radiation-free, accessible, and cost-effective postural analysis diagnostic tool after the EOS era.
Files in This Item:
T202301716.pdf Download
DOI
10.3390/healthcare11050686
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Rehabilitation Medicine (재활의학교실) > 1. Journal Papers
Yonsei Authors
Park, Jung Hyun(박중현) ORCID logo https://orcid.org/0000-0003-3262-7476
Cho, Han Eol(조한얼) ORCID logo https://orcid.org/0000-0001-5625-3013
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/193743