Synthesis of T2-weighted images from proton density images using a generative adversarial network in a temporomandibular joint magnetic resonance imaging protocol

Authors
 Lee, Chena  ;  Ha, Eun-Gyu  ;  Choi, Yoon joo  ;  Jeon, Kug Jin  ;  Han, Sang Sun 
Citation
 Imaging Science in Dentistry, Vol.52(4) : 393-398, 2022-12 
Journal Title
IMAGING SCIENCE IN DENTISTRY
ISSN
 2233-7822 
Issue Date
2022-12
Keywords
WORDS ; Deep Learning ; Computer Neural Network ; Artificial Intelligence ; Temporomandibular Joint Disorders ; Magnetic Resonance Imaging
Abstract
Purpose: This study proposed a generative adversarial network (GAN) model for T2-weighted image (WI) synthesis from proton density (PD)-WI in a temporomandibular joint (TMJ) magnetic resonance imaging (MRI) protocol. Materials and Methods: From January to November 2019, MRI scans for TMJ were reviewed and 308 imaging sets were collected. For training, 277 pairs of PD- and T2-WI sagittal TMJ images were used. Transfer learning of the pix2pix GAN model was utilized to generate T2-WI from PD-WI. Model performance was evaluated with the structural similarity index map (SSIM) and peak signal-to-noise ratio (PSNR) indices for 31 predicted T2-WI (pT2). The disc position was clinically diagnosed as anterior disc displacement with or without reduction, and joint effusion as present or absent. The true T2-WI-based diagnosis was regarded as the gold standard, to which pT2-based diagnoses were compared using Cohen's. coefficient. Results: The mean SSIM and PSNR values were 0.4781(+/- 0.0522) and 21.30(+/- 1.51) dB, respectively. The pT2 protocol showed almost perfect agreement (.= 0.81) with the gold standard for disc position. The number of discordant cases was higher for normal disc position (17%) than for anterior displacement with reduction (2%) or without reduction (10%). The effusion diagnosis also showed almost perfect agreement (.= 0.88), with higher concordance for the presence (85%) than for the absence (77%) of effusion. Conclusion: The application of pT2 images for a TMJ MRI protocol useful for diagnosis, although the image quality of pT2 was not fully satisfactory. Further research is expected to enhance pT2 quality.
DOI
10.5624/isd.20220125
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Oral and Maxillofacial Radiology (영상치의학교실) > 1. Journal Papers
Yonsei Authors
Lee, Chena(이채나) ORCID logo https://orcid.org/0000-0002-8943-4192
Jeon, Kug Jin(전국진) ORCID logo https://orcid.org/0000-0002-5862-2975
Choi, Yoon Joo(최윤주) ORCID logo https://orcid.org/0000-0001-9225-3889
Han, Sang Sun(한상선) ORCID logo https://orcid.org/0000-0003-1775-7862
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/193262
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links