Evaluating Mobius3D Dose Calculation Accuracy for Small-Field Flattening- Filter-Free Photon Beams

Authors
 Changhwan Kim  ;  Jihun Kim  ;  Young Kyu Lee  ;  Han-Back Shin  ;  Min Cheol Han  ;  Hojin Kim  ;  Jin Sung Kim 
Citation
 TECHNOLOGY IN CANCER RESEARCH & TREATMENT, Vol.21 : 15330338221141542, 2022 
Journal Title
TECHNOLOGY IN CANCER RESEARCH & TREATMENT
ISSN
 1533-0346 
Issue Date
2022
MeSH
Humans ; Particle Accelerators ; Radiometry / methods ; Radiosurgery* ; Radiotherapy Dosage ; Radiotherapy Planning, Computer-Assisted / methods ; Radiotherapy, Intensity-Modulated* / methods ; Retrospective Studies
Keywords
Mobius3D ; SBRT ; pretreatment quality assurance ; small field ; unflattened x-ray
Abstract
Purpose: We aimed to investigate the dose calculation accuracy of Mobius3D for small-field flattening-filter-free x-rays, mainly utilized for stereotactic body radiation therapy (SBRT). The accuracy of beam modeling and multileaf collimator (MLC) modeling in Mobius3D, significantly affecting the dose calculation is investigated. Methods: The commissioning procedures of Mobius3D were performed for unflattened 6 MV and 10 MV x-ray beams of the linear accelerator, including beam model adjustment and dosimetric leaf gap (DLG) optimization. An experimental study with artificial plans was conducted to evaluate the accuracy of small-field modeling. The dose calculation accuracy of Mobius3D was also evaluated for retrospective SBRT plans with multiple targets. Results: Both studies evaluated the dose calculation accuracy through comparisons with the measured data. Relatively large differences were observed for off-axis distances over 5 cm and for small fields less than 1 cm field size. For the study with artificial plans, the maximum absolute error of 9.96% for unflattened 6 MV and 9.07% for unflattened 10 MV was observed when the field size was 1 cm. For the study with patient plans, the mean gamma passing rate with 3%/3 mm gamma criterion was 63.6% for unflattened 6 MV and 82.6% for unflattened 10 MV. The maximum of the average dose difference was -19.9% for unflattened 6MV and -10.1% for unflattened 10MV. Conclusions: The dose calculation accuracy uncertainties of Mobius3D for small-field flattening-filter-free photon beams were observed. The study results indicated that the beam and MLC modeling of Mobius3D must be improved for use in SBRT pretreatment QA in clinical practice.
Files in This Item:
T202300110.pdf Download
DOI
10.1177/15330338221141542
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jihun(김지훈) ORCID logo https://orcid.org/0000-0003-4856-6305
Kim, Jinsung(김진성) ORCID logo https://orcid.org/0000-0003-1415-6471
Kim, Changhwan(김창환)
Kim, Hojin(김호진) ORCID logo https://orcid.org/0000-0002-4652-8682
Shin, Han-Back(신한백)
Han, Min Cheol(한민철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/192911
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links