126 225

Cited 0 times in

Potential Efficacy of Multiple-shot Long-pulsed 1,064-nm Nd:YAG in Nonablative Skin Rejuvenation: A Pilot Study

Authors
 Young-Koo Kim  ;  Hae-Jin Lee  ;  Jihee Kim 
Citation
 Medical Lasers, Vol.9(2) : 159-165, 2020-12 
Journal Title
Medical Lasers
ISSN
 2287-8300 
Issue Date
2020-12
Keywords
Nonablative skin rejuvenation ; Long-pulsed Nd:YAG laser ; Shot stacking ; Heat shock proteins, Integral IR thermal sensor
Abstract
Background and Objectives
The ultimate goal in current skin rejuvenation practice is to achieve a good result with minimal pain and downtime. Nonablative skin rejuvenation (NSR) is one technique. The efficacy of the long-pulsed 1064 nm Nd:YAG laser (LPNDY) has not been assessed in NSR.
Materials and Methods
Three target areas were selected (bilateral cheeks and glabellar region) in six volunteer subjects. A LPNDY with an integral skin temperature monitor delivered three stacked shots to each target area (1064 nm, 12 mm spot, 13 J/cm2, 1 Hz) without any skin cooling or anesthesia. The skin temperature was recorded before, during, and after each set of shots using the system monitor and in real-time using a high-sensitivity (±0.001°C) near-infrared video camera. The skin reaction was observed with the naked eye, and pain and discomfort were assessed by the subjects during and after treatment.
Results
The subjects reported a mild feeling of heat with no discomfort during or after the test treatments. Mild erythema was observed around the treatment areas, without noticeable edema. A series of three ascending skin temperature stepwise peaks, with a decrease in skin temperature towards the baseline after the third shot, was observed consistently. The mean temperatures for shots 1, 2, and 3 for the cheeks were 39.5°C, 42.0°C, and 44.4°C, respectively, and for the glabella, 40.8°C, 43.9°C, and 46.2°C, respectively. Similar ranges were indicated on the system integral temperature monitor.
Conclusion
A set of three stacked pulses with the LPNDY at a low fluence achieved ideal dermal temperatures to achieve some dermal remodeling but without any downtime or adverse events. The temperature data from the integral thermal sensor matched the video camera measurements with practical accuracy for skin rejuvenation requirements. These data suggest that LPNDY would satisfy the necessary criteria to achieve effective NSR, but further studies will be needed to assess the actual results in clinical practice.
Files in This Item:
T202007423.pdf Download
DOI
10.25289/ML.2020.9.2.159
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Dermatology (피부과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jihee(김지희) ORCID logo https://orcid.org/0000-0002-0047-5941
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/191145