62 198

Cited 8 times in

Errors according to the number of registered markers used in navigation-assisted surgery of the mandible

Authors
 Young-Eun Hwang 1, Sang-Hoon Kang 2 3, Hang-Keun Kim 4 5 
Citation
 HEAD & FACE MEDICINE, Vol.15(1) : 6, 2019-02 
Journal Title
HEAD & FACE MEDICINE
Issue Date
2019-02
MeSH
Fiducial Markers* ; Humans ; Imaging, Three-Dimensional ; Mandible* / surgery ; Surgery, Computer-Assisted*
Keywords
Navigation surgery ; Image-guided surgery ; Mandible ; Registration
Abstract
BackgroundThe aim of this study was to evaluate the accuracy of navigation according to the number of markers in terms of target registration errors (TREs) at each anatomical location during the registration process of the navigation system for the mandible.MethodsThe TREs were measured in five different experiments, varying only in the number of registration reference markers, which ranged from three to seven. To measure the TREs according to the number of registration reference markers, two experimental navigation devices were used: 1) Cbyon navigation surgery equipment 2) Polaris optical tracker. Both experiments were conducted to obtain the TREs at the anatomical locations of the mandible according to the number of registration markers during the navigation process. Statistical analysis was performed using the SPSS 23.0 software.ResultsAt all anatomical locations, errors were 2mm or less. Further, significant differences in the target errors measured by the Cbyon system were found according to the number of registration markers. Significant differences in the target errors measured by the Polaris optical tracker were found according to the registration markers at the posterior border only. In both groups, the target errors did not decrease as the number of registration markers increased.ConclusionsThis study demonstrates that an increase in the number of registration markers is not associated with a decrease in the TRE, and that a specific number of registration markers could reduce the TREs at each anatomical site. It is important to determine the minimum number of image registration markers at which the smallest TRE would be observed for different surgical sites.
Files in This Item:
T9992019145.pdf Download
DOI
10.1186/s13005-019-0190-z
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Oral and Maxillofacial Surgery (구강악안면외과학교실) > 1. Journal Papers
Yonsei Authors
Kang, Sang Hoon(강상훈)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/189148
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links