355 510

Cited 0 times in

Accuracy of the Kane Formula for Intraocular Lens Power Calculation in Comparison with Existing Formulas: A Retrospective Review

Authors
 Soyoung Ryu  ;  Ikhyun Jun  ;  Tae-Im Kim  ;  Eung Kweon Kim  ;  Kyoung Yul Seo 
Citation
 YONSEI MEDICAL JOURNAL, Vol.62(12) : 1117-1124, 2021-12 
Journal Title
YONSEI MEDICAL JOURNAL
ISSN
 0513-5796 
Issue Date
2021-12
MeSH
Biometry ; Humans ; Lens Implantation, Intraocular ; Lenses, Intraocular* ; Optics and Photonics ; Phacoemulsification* ; Refraction, Ocular ; Retrospective Studies
Keywords
Barrett Universal II ; IOL power calculation ; Kane ; refractive outcomes
Abstract
Purpose: To evaluate the accuracy of the Kane formula for intraocular lens (IOL) power calculation in comparison with existing formulas by incorporating optional variables into calculation.

Materials and methods: This retrospective review consisted of 78 eyes of patients who had undergone uneventful phacoemulsification with intraocular implantation at Severance Hospital in Seoul, Korea between February 2020 and January 2021. The Kane formula was compared with six of the existing IOL formulas (SRK/T, Hoffer-Q, Haigis, Holladay1, Holladay2, Barrett Universal II) based on the mean absolute error (MAE), median absolute error (MedAE), and the percentages of eyes within prediction errors of ±0.25D, ±0.50D, and ±1.00D.

Results: The Barrett Universal II formula demonstrated the lowest MAEs (0.26±0.17D), MedAEs (0.28D), and percentage of eyes within prediction errors of ±0.25D, ± 0.50D, and ±1.00D, although there was no statistically significant difference between Barrett Universal II-SRK/T (p=0.06), and Barrett Universal II-Kane formula (p<0.51). Following the Barrett Universal II formula, the Kane formula demonstrated the second most accurate formula with MAEs (0.30±0.19D) and MedAEs (0.28D). However, no statistical difference was shown between Kane-Barrett Universal II (p=0.51) and Kane-SRK/T (p=0.14).

Conclusion: Although slightly better refractory outcome was noted in the Barrett Universal II formula, the performance of the Kane formula in refractive prediction was comparable in IOL power calculation, marking its superiority over many conventional IOL formulas, such as HofferQ, Haigis, Holladay1, and Holladay2.
Files in This Item:
T202105562.pdf Download
DOI
10.3349/ymj.2021.62.12.1117
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Jun, Ik Hyun(전익현) ORCID logo https://orcid.org/0000-0002-2160-1679
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/187319
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.