Cited 0 times in

Comparison of Retrievability and Indwelling Complications of Celect and Denali Infrarenal Vena Cava Filters: A Randomized, Controlled Trial

DC Field Value Language
dc.contributor.author권준호-
dc.contributor.author김경민-
dc.contributor.author김만득-
dc.contributor.author문성모-
dc.contributor.author원종윤-
dc.contributor.author한기창-
dc.contributor.author이준형-
dc.date.accessioned2021-12-28T16:55:42Z-
dc.date.available2021-12-28T16:55:42Z-
dc.date.issued2021-10-
dc.identifier.issn0174-1551-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/186858-
dc.description.abstractPurpose: To compare the Celect and Denali filters in terms of complex filter retrieval and indwelling complications after a 2-month indwelling time. Materials and methods: In this prospective, randomized trial, 153 subjects were assessed for eligibility between May 2016 and July 2018. A total of 136 participants were randomly assigned to receive either Celect (n = 68) or Denali (n = 68) filter placement in the infrarenal inferior vena cava. Tilt angles at placement and retrieval and rates of overall filter retrieval, indwelling complication, and complex retrieval were compared. Results: Of 136 participants (mean age, 62 ± 12.8 years, 55 male), 24 (17.6%) were lost to follow-up. The mean indwelling time of filter was 60.4 ± 7 days and there was no significant difference in the baseline characteristics between the two groups. Filter retrieval was successful in all participants (112/112, 100%). Significantly higher rates of filter tilt > 15° (n = 8) and strut penetration (n = 14) were found with the Celect filter than with the Denali filter (1 significant tilt and 1 penetration) (P = 0.033 and 0.001, respectively). No filter fractures were observed and there was no significant difference in tip embedment, filter fracture, filter migration, or mean fluoroscopy times. There were 3 cases of complex retrieval (1 for Denali vs. 2 for Celect, P = 0.500), for which the loop-snare technique was used. Conclusion: Denali filters demonstrated significantly lower rates of tilt angle > 15° and strut penetration. However, there was no significant difference in the complex filter retrieval rate between the Celect and Denali filters.-
dc.description.statementOfResponsibilityrestriction-
dc.languageEnglish-
dc.publisherSpringer Verlag-
dc.relation.isPartOfCARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.subject.MESHAged-
dc.subject.MESHDevice Removal-
dc.subject.MESHHumans-
dc.subject.MESHMale-
dc.subject.MESHMiddle Aged-
dc.subject.MESHProspective Studies-
dc.subject.MESHProsthesis Design-
dc.subject.MESHPulmonary Embolism*-
dc.subject.MESHRetrospective Studies-
dc.subject.MESHTreatment Outcome-
dc.subject.MESHVena Cava Filters* / adverse effects-
dc.subject.MESHVena Cava, Inferior / diagnostic imaging-
dc.titleComparison of Retrievability and Indwelling Complications of Celect and Denali Infrarenal Vena Cava Filters: A Randomized, Controlled Trial-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiology (영상의학교실)-
dc.contributor.googleauthorKichang Han-
dc.contributor.googleauthorMan-Deuk Kim-
dc.contributor.googleauthorGyoung Min Kim-
dc.contributor.googleauthorJoon Ho Kwon-
dc.contributor.googleauthorJunhyung Lee-
dc.contributor.googleauthorJong Yun Won-
dc.contributor.googleauthorSungmo Moon-
dc.contributor.googleauthorHyeongsu An-
dc.identifier.doi10.1007/s00270-021-02928-9-
dc.contributor.localIdA05085-
dc.contributor.localIdA00296-
dc.contributor.localIdA00420-
dc.contributor.localIdA06146-
dc.contributor.localIdA02443-
dc.contributor.localIdA05062-
dc.relation.journalcodeJ00459-
dc.identifier.eissn1432-086X-
dc.identifier.pmid34312689-
dc.identifier.urlhttps://link.springer.com/article/10.1007%2Fs00270-021-02928-9-
dc.subject.keywordDeep venous thrombosis-
dc.subject.keywordFilter-
dc.subject.keywordInferior vena cava-
dc.subject.keywordPulmonary embolism-
dc.subject.keywordRetrieval-
dc.contributor.alternativeNameKwon, Joon Ho-
dc.contributor.affiliatedAuthor권준호-
dc.contributor.affiliatedAuthor김경민-
dc.contributor.affiliatedAuthor김만득-
dc.contributor.affiliatedAuthor문성모-
dc.contributor.affiliatedAuthor원종윤-
dc.contributor.affiliatedAuthor한기창-
dc.citation.volume44-
dc.citation.number10-
dc.citation.startPage1536-
dc.citation.endPage1542-
dc.identifier.bibliographicCitationCARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, Vol.44(10) : 1536-1542, 2021-10-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers

qrcode