The impedance pneumography which is widely used in monitoring respiration is simple to use and noninvasive, but it is sensitive to motion artifacts and insensitive to detect obstructive apnea. A 3-channel respiratory inductive plethysmography (RIP) developed in this study detects inductance change of the inductance band induced by cross-sectional area change of thorax or abdomen as one breathes. It was confirmed that RIP was less sensitive to various motion artifacts but more sensitive to detection of obstructive apnea than impedance pneumography.