0 335

Cited 0 times in

Performance evaluation of the Roche cobas 6800 system for quantifying cytomegalovirus DNA in plasma and urine samples

Authors
 Juhye Roh  ;  Sinyoung Kim  ;  Eunmin Kwak  ;  Jungyong Park  ;  Younhee Park 
Citation
 JOURNAL OF CLINICAL VIROLOGY, Vol.138 : 104816, 2021-05 
Journal Title
JOURNAL OF CLINICAL VIROLOGY
ISSN
 1386-6532 
Issue Date
2021-05
Keywords
Cytomegalovirus ; Nucleic acid amplification testing ; Real-time polymerase chain reaction ; Urine sample ; Viral load
Abstract
Introduction: Cytomegalovirus (CMV) nucleic acid amplification testing is important for CMV infection diagnosis and management. CMV DNA is found in plasma and various other fluids, including urine. If CMV can be reliably detected in urine, it may be considered a non-invasive alternative to blood tests. The cobas 6800 system (Roche Diagnostics, Mannheim, Germany) is a Food and Drug Administration-approved testing platform for measuring CMV DNA in plasma.

Objective: To evaluate the analytical performance of the cobas 6800 system and compare the clinical feasibility of CMV detection in plasma and urine samples.

Study design: Imprecision, linearity, limit of quantitation (LOQ), and cross-reactivity of the cobas 6800 system were assessed, and reference interval verification was performed. Plasma CMV DNA quantification was compared to CMV DNA values in urine samples obtained from 129 pediatric patients (<18 years of age) from March 2020 to May 2020 at a tertiary hospital.

Results: The assay precision was within the acceptable range. Linearity was observed within the tested concentration range (2.36-6.33 log IU/mL) with a coefficient of determination of 0.9972. The LOQ was 34.5 IU/mL. The assay did not show cross-reactivity with 15 other viruses. Plasma and urine detection results were stratified into three categories: negative,
Conclusion: The cobas 6800 system offers good sensitivity, precision, and linearity and is suitable for monitoring CMV viral loads in the plasma and urine samples.
Full Text
https://www.sciencedirect.com/science/article/pii/S1386653221000834
DOI
10.1016/j.jcv.2021.104816
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Sin Young(김신영) ORCID logo https://orcid.org/0000-0002-2609-8945
Park, Youn Hee(박윤희) ORCID logo https://orcid.org/0000-0001-8458-1495
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/184059
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links