Flexible film dosimeter for in vivo dosimetry

Authors
 Jin Dong Cho  ;  Jaeman Son  ;  Jiwon Sung  ;  Chang Heon Choi  ;  Jin Sung Kim  ;  Hong-Gyun Wu  ;  Jong Min Park  ;  Jung-In Kim 
Citation
 MEDICAL PHYSICS, Vol.47(7) : 3204-3213, 2020-07 
Journal Title
MEDICAL PHYSICS
ISSN
 0094-2405 
Issue Date
2020-07
Keywords
flexible dosimeter ; in vivo dosimetry ; radiochromic film
Abstract
Purpose: The aims of this study were to develop a flexible film dosimeter applicable to the irregular surface of a patient for in vivo dosimetry and to evaluate the device's dosimetric characteristics.

Methods: A flexible film dosimeter with active layers consisting of radiochromic-sensitive films and flexible silicone materials was constructed. The dose-response, sensitivity, scanning orientation dependence, energy dependence, and dose rate dependence of the flexible film dosimeter were tested. Irradiated dosimeters were scanned 24 h post-irradiation, and the region of interest was 5 mm × 5 mm. Biological stability tests ensured the safety of application of the flexible film dosimeter for patients. A preliminary clinical study with the flexible film dosimeter was implemented on four patients.

Results: The red channel demonstrated the highest sensitivity among all channels, and the response sensitivity of the dosimeter decreased with the applied dose, which were the same as the characteristics of GAFCHROMIC EBT3 radiochromic films. The flexible film dosimeter showed no significant energy dependence for photon beams of 6 MV, 6 MV flattening filter-free (FFF), 10 MV, and 15 MV. The flexible film dosimeter showed no substantial dose rate dependence with 6 or 6 MV FFF. In terms of biological stability, the flexible film dosimeter demonstrated no cytotoxicity, no irritation, and no skin sensitization. In the preliminary clinical study, the dose differences between the measurements with the flexible film dosimeter and calculations with the treatment planning system ranged from -0.1% to 1.2% for all patients.

Conclusions: The dosimeter developed in this study is a flexible film capable of attachment to a curved skin surface. The biological test results indicate the stability of the flexible film dosimeter. The preliminary clinical study showed that the flexible film dosimeter can be successfully applied as an in vivo dosimeter.
Full Text
https://aapm.onlinelibrary.wiley.com/doi/full/10.1002/mp.14162
DOI
10.1002/mp.14162
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jinsung(김진성) ORCID logo https://orcid.org/0000-0003-1415-6471
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/180292
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links