Abstract
Purpose: To investigate the efficacy and safety of a newly developed pinhole soft contact lens (Eyelike Pinhole II; Koryo Eyetech Co. Ltd.) for presbyopia correction.

Methods: This prospective clinical study enrolled 29 patients with presbyopia between October 2018 and December 2018. All participants wore the Eyelike Pinhole II in the non-dominant eye for >3 h/day for a period of 1 week. Binocular and monocular uncorrected near visual acuities, distance-corrected near visual acuity (DCNVA), uncorrected distance visual acuity, and corrected distance visual acuity (CDVA) were measured before and after the intervention. All visual acuities were measured in logarithm of the minimal angle of resolution (logMAR) units. In addition, binocular defocus curves were generated, and contrast sensitivity values were obtained under photopic and mesopic conditions before and after lens wear.

Results: The mean DCNVA of the treated eye and the mean binocular DCNVA improved from 0.34 ± 0.12 to 0.15 ± 0.14 (P < 0.001) and 0.31 ± 0.13 to 0.11 ± 0.10 (P < 0.001) logMAR, respectively, after pinhole contact lens wear. Although the mean CDVA of the treated eye deteriorated from -0.04 ± 0.05 to 0.02 ± 0.11 logMAR (P = 0.015), there was no significant change in the mean binocular CDVA (P = 0.79). The binocular defocus curve showed a significant improvement from -5.0 dioptres (D) to -1.0 D after pinhole contact lens wear.

Conclusions: The newly developed Eyelike Pinhole II soft contact lens showed safe and effective outcomes; thus, it could be a promising option for the treatment of presbyopia.
Full Text
https://www.sciencedirect.com/science/article/pii/S1367048419302711
DOI
10.1016/j.clae.2019.12.009
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
Seo, Kyoung Yul(서경률) ORCID logo https://orcid.org/0000-0002-9855-1980
Jun, Ik Hyun(전익현) ORCID logo https://orcid.org/0000-0002-2160-1679
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/180073
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links