0 266

Cited 4 times in

Development and evaluation of an automated stainer for acid-fast bacilli

Authors
 Soo Chan Kim  ;  Seung Il Kang  ;  Deok Won Kim  ;  Seung Cheol Kim  ;  Sang-Nae Cho  ;  Jung Ho Hwang  ;  Young Kim  ;  Sun-Dae Song  ;  Young Ha Kim 
Citation
 MEDICAL ENGINEERING & PHYSICS, Vol.25(4) : 341-347, 2003-05 
Journal Title
MEDICAL ENGINEERING & PHYSICS
ISSN
 1350-4533 
Issue Date
2003-05
MeSH
Equipment Design ; Hot Temperature ; Humans ; Mycobacterium tuberculosis / ultrastructure* ; Robotics / instrumentation* ; Robotics / methods ; Rosaniline Dyes ; Single-Blind Method ; Specimen Handling / instrumentation* ; Specimen Handling / methods ; Sputum / microbiology* ; Staining and Labeling / instrumentation* ; Staining and Labeling / methods ; Tissue Fixation / instrumentation ; Tissue Fixation / methods ; Tuberculosis, Pulmonary / pathology*
Abstract
The current strategy for the control of tuberculosis (TB) relies on early diagnosis, and smear microscopy is an essential component of the laboratory diagnosis of TB in most countries with a high prevalence of the disease. However, even simple smear microscopy examination is far from satisfactory because staining results can vary among individual technicians. In an effort to minimize variations in manual staining procedures, we developed an automated stainer for AFB and evaluated its usefulness in comparison with manual staining. The key feature of our automated stainer is a heating apparatus required for fixation and carbol-fuchsin staining. After smear slides are placed into the machine, the entire staining process is fully automated, from fixation to final washing and drying. With the automated methods, five slides can be fixed and stained in 21 min at consistent high quality. Using sputum samples from 91 TB patients, the staining results of the automated stainer were compared blindly with those of manual staining. The concordance rate between the two methods was 94.5%. In addition, there was no significant difference in the rate of detection of AFB in the sputum samples. Although further optimization of the auto staining procedures is required, the results indicate that the automated AFB stainer developed in this study looks promising for use in clinical mycobacteriology laboratory in order to minimize personal variation during AFB staining.
Full Text
https://www.sciencedirect.com/science/article/pii/S1350453302002138
DOI
10.1016/s1350-4533(02)00213-8
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Engineering (의학공학교실) > 1. Journal Papers
Yonsei Authors
Kim, Deok Won(김덕원)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/178767
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links