A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery.
The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable
hemodynamics and underlying systolic left ventricular dysfunction. ICD interrogation revealed recurrent episodes of ICD
shock due to noise sensing and increased impedance of right ventricular (RV)-lead. With the impression of lead fracture,
ICD lead extraction was performed. The fractured ICD lead was completely removed by traction of locking stylet and countertraction
of polypropylene dilator sheath. A new lead was inserted and the patient was discharged without complications after
2 days. To our knowledge, this is the first report on ICD lead extraction by conventional traction and counter-traction technique
in Korea.
Files in This Item:
T201102539.pdf Download
DOI
10.4070 / kcj.2010.40.8.418
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
Yonsei Authors
Pak, Hui Nam(박희남) ORCID logo https://orcid.org/0000-0002-3256-3620
Lee, Moon-Hyoung(이문형) ORCID logo https://orcid.org/0000-0002-7268-0741
Joung, Bo Young(정보영) ORCID logo https://orcid.org/0000-0001-9036-7225
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/178656
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links