Implantable Cardioverter-Defibrillator Lead Extraction by Conventional Traction and Counter-Traction Technique
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 박희남 | - |
dc.contributor.author | 이문형 | - |
dc.contributor.author | 정보영 | - |
dc.date.accessioned | 2020-08-14T03:15:50Z | - |
dc.date.available | 2020-08-14T03:15:50Z | - |
dc.date.issued | 2010-08 | - |
dc.identifier.issn | 1738-5520 | - |
dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/178656 | - |
dc.description.abstract | A 46-year-old man presented to our institution with inappropriate implantable cardioverter-defibrillator (ICD) shock delivery. The ICD (single chamber, dual shock coils) was implanted for sustained monomorphic ventricular tachycardia with unstable hemodynamics and underlying systolic left ventricular dysfunction. ICD interrogation revealed recurrent episodes of ICD shock due to noise sensing and increased impedance of right ventricular (RV)-lead. With the impression of lead fracture, ICD lead extraction was performed. The fractured ICD lead was completely removed by traction of locking stylet and countertraction of polypropylene dilator sheath. A new lead was inserted and the patient was discharged without complications after 2 days. To our knowledge, this is the first report on ICD lead extraction by conventional traction and counter-traction technique in Korea. | - |
dc.description.statementOfResponsibility | open | - |
dc.language | English, Korean | - |
dc.publisher | Korean Society of Circulation | - |
dc.relation.isPartOf | Korean Circulation Journal | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.title | Implantable Cardioverter-Defibrillator Lead Extraction by Conventional Traction and Counter-Traction Technique | - |
dc.type | Article | - |
dc.contributor.college | College of Medicine (의과대학) | - |
dc.contributor.department | Dept. of Internal Medicine (내과학교실) | - |
dc.contributor.googleauthor | Jong Sung Park | - |
dc.contributor.googleauthor | Hui-Nam Pak | - |
dc.contributor.googleauthor | Moon-Hyoung Lee | - |
dc.contributor.googleauthor | Sung Soon Kim | - |
dc.contributor.googleauthor | Boyoung Joung | - |
dc.identifier.doi | 10.4070 / kcj.2010.40.8.418 | - |
dc.contributor.localId | A01776 | - |
dc.contributor.localId | A02766 | - |
dc.contributor.localId | A03609 | - |
dc.relation.journalcode | J01952 | - |
dc.identifier.eissn | 1738-5555 | - |
dc.subject.keyword | Device failure | - |
dc.subject.keyword | Implantable cardioverter defibrillator | - |
dc.subject.keyword | Device removal | - |
dc.contributor.alternativeName | Pak, Hui Nam | - |
dc.contributor.affiliatedAuthor | 박희남 | - |
dc.contributor.affiliatedAuthor | 이문형 | - |
dc.contributor.affiliatedAuthor | 정보영 | - |
dc.citation.volume | 40 | - |
dc.citation.number | 3 | - |
dc.citation.startPage | 418 | - |
dc.citation.endPage | 420 | - |
dc.identifier.bibliographicCitation | Korean Circulation Journal, Vol.40(3) : 418-420, 2010-08 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.