174 389

Cited 0 times in

심전도 전극에 의한 화상

Other Titles
 Electrosurgical Burns at EKG Electrode -A case report- 
Authors
 이성진  ;  강정완  ;  김원옥 
Citation
 Journal of Korean Society of Anesthesiologist (대한마취과학회지), Vol.34(2) : 461-463, 1998 
Journal Title
Journal of Korean Society of Anesthesiologist(대한마취과학회지)
ISSN
 0302-5780 
Issue Date
1998
Abstract
Electrosurgery can pose some risk to patients. Recently we experienced a case of accidental burning. Nine year-old patient received burn at the site of the ECG electrode attached to the right subclavian area during an electrosurgical operation. Even though the cutting and coagulation powers were not so much high, inappropriate earth with the fault of electrosurgical unit might lead to the generation of heat. To avoid recurrence, every electrical instruments should inspected, tested and required to be fault-free. Properly functioning surgical and monitoring units must be used.