174 389

Cited 0 times in

심전도 전극에 의한 화상

DC Field Value Language
dc.contributor.author김원옥-
dc.date.accessioned2020-07-02T17:32:09Z-
dc.date.available2020-07-02T17:32:09Z-
dc.date.issued1998-
dc.identifier.issn0302-5780-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/176875-
dc.description.abstractElectrosurgery can pose some risk to patients. Recently we experienced a case of accidental burning. Nine year-old patient received burn at the site of the ECG electrode attached to the right subclavian area during an electrosurgical operation. Even though the cutting and coagulation powers were not so much high, inappropriate earth with the fault of electrosurgical unit might lead to the generation of heat. To avoid recurrence, every electrical instruments should inspected, tested and required to be fault-free. Properly functioning surgical and monitoring units must be used.-
dc.description.statementOfResponsibilityopen-
dc.languageKorean-
dc.publisher대한마취과학회-
dc.relation.isPartOfJournal of Korean Society of Anesthesiologist (대한마취과학회지)-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.title심전도 전극에 의한 화상-
dc.title.alternativeElectrosurgical Burns at EKG Electrode -A case report--
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Anesthesiology and Pain Medicine (마취통증의학교실)-
dc.contributor.googleauthor이성진-
dc.contributor.googleauthor강정완-
dc.contributor.googleauthor김원옥-
dc.contributor.localIdA00766-
dc.relation.journalcodeJ01540-
dc.contributor.alternativeNameKim, Won Oak-
dc.contributor.affiliatedAuthor김원옥-
dc.citation.volume34-
dc.citation.number2-
dc.citation.startPage461-
dc.citation.endPage463-
dc.identifier.bibliographicCitationJournal of Korean Society of Anesthesiologist (대한마취과학회지), Vol.34(2) : 461-463, 1998-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.