Balloon-Supported Passage of a Stent-Graft into the Aortic Arch

DC Field Value Language
dc.contributor.author송석원-
dc.contributor.author은나래-
dc.contributor.author이광훈-
dc.date.accessioned2018-11-01T16:40:47Z-
dc.date.available2018-11-01T16:40:47Z-
dc.date.issued2015-
dc.identifier.issn1229-6929-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/164982-
dc.description.abstractA 62-year-old man was admitted, and thoracic endovascular aortic repair (TEVAR) procedure was performed to treat an accidentally detected aortic aneurysm, which was 63 mm in diameter. While performing TEVAR, the passage of the stent-graft introducer system was impossible due to the prolapse of the introducer system into a wide-necked aneurysm; this aneurysm was located at the greater curvature of the proximal descending thoracic aorta. In order to advance the introducer system, a compliant balloon was inflated. Thus, we created an artificial wall in the aneurysm with this inflated balloon. Finally, we were able to advance the introducer system into the target zone.-
dc.description.statementOfResponsibilityopen-
dc.formatapplication/pdf-
dc.languageEnglish-
dc.publisherKorean Society of Radiology-
dc.relation.isPartOfKOREAN JOURNAL OF RADIOLOGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rightshttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.titleBalloon-Supported Passage of a Stent-Graft into the Aortic Arch-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Thoracic and Cardiovascular Surgery (흉부외과학교실)-
dc.contributor.googleauthorNa Lae Eun-
dc.contributor.googleauthorDahye Lee-
dc.contributor.googleauthorSuk-Won Song-
dc.contributor.googleauthorSeung-Moon Joo-
dc.contributor.googleauthorTilo Kölbel-
dc.contributor.googleauthorKwang-Hun Lee-
dc.identifier.doi10.3348/kjr.2015.16.4.744-
dc.contributor.localIdA02028-
dc.contributor.localIdA04778-
dc.contributor.localIdA02676-
dc.relation.journalcodeJ02884-
dc.identifier.eissn2005-8330-
dc.identifier.pmid26175573-
dc.subject.keywordAortic arch-
dc.subject.keywordBalloon-supported passage of an introducer system-
dc.subject.keywordProximal descending thoracic aorta-
dc.subject.keywordThoracic aortic aneurysm-
dc.subject.keywordThoracic endovascular aortic repair-
dc.subject.keywordWide-necked aneurysm-
dc.contributor.alternativeNameSong, Suk Won-
dc.contributor.alternativeNameEun, Na Lae-
dc.contributor.alternativeNameLee, Kwang Hun-
dc.contributor.affiliatedAuthor송석원-
dc.contributor.affiliatedAuthor은나래-
dc.contributor.affiliatedAuthor이광훈-
dc.citation.volume16-
dc.citation.number4-
dc.citation.startPage744-
dc.citation.endPage748-
dc.identifier.bibliographicCitationKOREAN JOURNAL OF RADIOLOGY, Vol.16(4) : 744-748, 2015-
dc.identifier.rimsid58305-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Thoracic and Cardiovascular Surgery (흉부외과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.